Xylem Lab 960 User Manual

Page 28

Advertising
background image

Operation

Lab 960

28

ba75557e05

11/2009

Calibration evaluation

After the calibration, the meter automatically evaluates the current
status. The evaluation appears on the display.

Calibration record

When finishing a calibration, the new calibration values are stored. )

Note
The calibration record is automatically transmitted to the interface after
calibrating.

Sample record:

Downloading calibration

data

You can download the calibration data:

z to the display (see page 43)

z to the interface (see page 45)

Display

Cell constant [cm

-1

]

Status display indicator CAL.

You are working with a
correctly calibrated measuring
cell.

in the range
0.450 ... 0.500 cm

-1

0.585 ... 0.715 cm

-1

0.800 ... 1.200 cm

-1

E3

Eliminate the error according to

CHAPTER

6 W

HAT

TO

DO

IF

...

(

PAGE

61).

outside the ranges
0.450 ... 0.500 cm

-1

0.585 ... 0.715 cm

-1

or
0.800 ... 1.200 cm

-1

16.09.2005 08:53:54
Lab 960 02320025

CALIBRATION COND
Cal Time : 16.09.2005 08:22:14
Cal Interval: 180d
Cal Std.: 0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const : 0.975 1/cm
Probe : +++