4.9
Reset
4.9.1
Resetting the cell constant
This function serves to erase the last determined cell constant.
Subsequently, the meter uses the last manually adjusted cell constant.
All other meter settings are retained.
Note
The measuring system is possibly not calibrated after a reset. Before
measuring, make sure the meter uses the cell constant suitable for the
measuring cell.
Resetting the cell
constant
1
Press <On/Off
_
> to open the menu for the reset of the cell
constant.
Ini.C is displayed.
2
Press <
▲><▼> to display no or YES.
YES: Reset the cell constant.
no: Retain the cell constant.
3
Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.