9 reset, 1 resetting the cell constant, Reset 4.9.1 – Xylem Lab 960 User Manual

Page 55: Resetting the cell constant, Ci n i

Advertising
background image

Lab 960

Operation

55

ba75557e05

11/2009

4.9

Reset

4.9.1

Resetting the cell constant

This function serves to erase the last determined cell constant.
Subsequently, the meter uses the last manually adjusted cell constant.
All other meter settings are retained.

Note
The measuring system is possibly not calibrated after a reset. Before
measuring, make sure the meter uses the cell constant suitable for the
measuring cell.

Resetting the cell

constant

1

Press <On/Off

_

> to open the menu for the reset of the cell

constant.
Ini.C is displayed.

2

Press <

▲><▼> to display no or YES.

YES: Reset the cell constant.
no: Retain the cell constant.

3

Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.

o

n

C

i

n

I

Advertising