Standard solution

The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.

TDS

Total dissolved solids
Calculation: TDS (mg/l) = conductivity (µS) * TDS factor (mg/l*µS)

Temperature

coefficient

Value of the slope

of a linear temperature function.

Temperature

compensation

Name of a function that considers the temperature influence on the
measurement and converts it accordingly. Depending on the
measured parameter to be determined, the temperature
compensation functions in different ways. For conductimetric
measurements, the measured value is converted to a defined
reference temperature. For potentiometric measurements, the slope
value is adjusted to the temperature of the test sample but the
measured value is not converted.

Temperature function

Name of a mathematical function expressing the temperature
behavior of a test sample, a sensor or part of a sensor.

Test sample

Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.

α

T

Ref

=

Meas*

1

1 + a

(T -

)

*

T

Ref