YSI Professional Plus User Manual

Page 16

Advertising
background image

22

23

SEtUP - DISSoLVED oXYGEN

Press Sensor

, highlight Setup and press enter. Next, highlight DO and press

enter.

Enabled allows you to enable or disable the

Dissolved oxygen function. Highlight Enabled
and press enter to activate(

Q

) or deactivate(

P

)

dissolved oxygen. Disable dissolved oxygen

if you do not have a dissolved oxygen sensor

connected to the instrument.

If a sensor is Enabled that isn’t connected to the

instrument, the display will show an unstable, false

reading, ?????, or ----- next to the units.

Sensor Type sets the type of oxygen sensor being used: either Polarographic

(black) or Galvanic (grey). Highlight Sensor Type and press enter. Highlight the

correct sensor type installed on the cable and press enter to confirm.

If using a ProBoD sensor/cable assembly, the sensor type should be set to

polarographic.

the Pro Plus has two compatible sensors for use with a field cable:

Polarographic – this sensor has a black sensor body and is engraved with the

model number 2003.

Galvanic – this sensor has a grey sensor body and is engraved with the model

number 2002.

In terms of physical configuration, membrane material, and general performance,

YSI Professional Series Galvanic dissolved oxygen sensors are exactly like the

Professional Series Polarographic sensors. the advantage of using Galvanic

sensors is convenience. Galvanic sensors provide for an instant-on sensor without

the need for warm-up time but this affects the life of the sensor. Polarographic

sensors last longer and have a longer warranty but require a 5-15 minute warm-

up time before use or calibration.

IMPORTANT – The instrument default setting is Galvanic.

Please change the Sensor Type to match the correct sensor. If

you observe readings very close to 0 or extremely high readings

(i.e. 600%), your Sensor Type setting (Polarographic or

Galvanic) may be set incorrectly and you should immediately

ensure it matches the sensor installed on your cable.

Membrane sets the type of membrane used on the dissolved oxygen sensor.

Highlight Membrane and press enter. Highlight the correct membrane type

installed on the sensor and press enter to confirm. the Do sensor is supplied

with membranes specific to the sensor type ordered and are color coded as

described in the following tables.

Galvanic membrane kits:
Item

Color

Material

Description

5912

Black

1 mil teflon®

traditional membrane material

5913

Yellow 1.25 mil polyethylene Improved response time and less

flow dependence than teflon®

Ships standard with the sensor.

5914

Blue

2 mil polyethylene

Less flow dependence than 1.25 mil

but somewhat slower response

Polarographic membrane kits:
Item

Color

Material

Description

5906

Black

1 mil teflon®

traditional membrane material

5908

Yellow 1.25 mil polyethylene Improved response time and less

flow dependence than teflon®

Ships standard with the sensor.

5909

Blue

2 mil polyethylene

Less flow dependence than 1.25 mil

but somewhat slower response

Selecting a Dissolved oxygen Membrane:

Membrane Type

Flow Dependence After 4 Minutes

Typical Response

Time - 95%

5912, 5906 - Black

60%

18 seconds

5913, 5908 - Yellow

25%

8 seconds

5914, 5909 - Blue

18%

17 seconds