YSI Professional Plus User Manual

Page 7

Advertising
background image

4

5

1010 dual cable

Port 1 Options

Port 2 Options

pH

pH

oRP

oRP

pH or pH/oRP*

pH or pH/oRP*

ammonium

ammonium

chloride

chloride

nitrate

nitrate
none (port plug)

* If using a 6051010 cable, a sensor must be installed in port 1 for correct

operation. If you install a pH/oRP combo sensor into a 6051010 cable, oRP

will not be measured. It is not recommended to use a pH/oRP combo sensor

on a 6051010 cable.

1020 dual cable

Port 1 Options

Port 2 Options

pH

Polarographic Do

oRP

Galvanic Do

pH or pH/oRP

none (port plug)

ammonium
chloride
nitrate
none (port plug)

If using a 1020 cable, install a pH, oRP, pH/oRP, Ammonium, Nitrate or

Chloride sensor in port 1 and a Do sensor in port 2.

If using a 605103 pH/ORP combination probe on a

6051020 or 6051030 cable you can report both pH

and ORP. However, it is recommended to set ISE1 as

pH and ISE2 as ORP in the Sensor Setup menu.

the Quatro cable bulkheads are labeled

1, 2, Do, and Ct, see figure to the left. All

sensors except the Conductivity/temperature

sensor can be installed following the Standard

Pro Series Sensor Installation instructions.

Conductivity/temperature sensor installation

is described in the next section. For ease of

installation, YSI recommends that you install

a sensor into port 1 first; followed by Do

installation, then port 2, and lastly C/t.

Quatro

Cable (pn:

605790)

Port 1

Options

Port 2

Options

DO Port

Options

CT Port Options

pH

pH

Polarographic

Do

5560

Conductivity/

temperature

sensor only

oRP

oRP

Galvanic Do

pH or pH/

oRP*

pH or pH/

oRP*

none (port

plug)

ammonium

ammonium

chloride

chloride

nitrate

nitrate
none (port

plug)

* If using a Quatro cable, a sensor must be installed in port 1 for correct

operation of port 2. If you install a pH/oRP combo sensor into a Quatro

cable, oRP will not be measured. It is not recommended to use a pH/oRP

combo sensor on a Quatro cable.

Before installing either dissolved oxygen sensor, the instrument

must be configured for the sensor being installed. See

the Setup - Dissolved Oxygen section of this manual for

instrument configuration instructions. Failure to do this

may result in damage not covered under warranty.

First, ensure both the sensor connector and sensor port on the cable are

clean and dry. to connect the sensor, grasp the sensor with one hand and the

sensor connection end of the cable (bulkhead) in the other. Push the sensor

into the connector on the cable until it is properly seated and only one o-ring

is visible. Failure to properly seat the probe may result in damage. twist the

sensor clockwise to engage threads and finger tighten (Figure 2). Do not use a

tool. this connection is waterproof. Please refer to the sensor installation sheet

that is included with each sensor for detailed instructions.