YSI Professional Plus User Manual

Page 18

Advertising
background image

26

27

Auto Stable and Hold All Readings enabled,

then the display will hold the readings once Do

and pH have both reached their Auto Stable

settings. You must press the Esc key to “release”

the held display in order to take subsequent

readings Hold All Readings must be reactivated

after each use!

SALINItY CoRRECtIoN

the last feature in the Sensor menu is the Salinity

correction value which is used to calculate the

dissolved oxygen mg/L and ammonia readings

when a conductivity sensor is not in use. Press
Sensor

, highlight Salinity, and press enter.

then, use the numeric entry screen to enter the

Salinity value of the water you will be testing

from 0 to 70 ppt.

If using a cable with a conductivity sensor, the salinity

measured by the conductivity sensor will be used in the Do and

ammonia mg/L calculations and ‘As Measured’ will be displayed next to Salinity

in the Sensor menu.

As the salinity of water increases, its ability to dissolve oxygen decreases. For

example, fully oxygenated 20 °C water at sea level with zero salinity will hold

9.092 mg/L of dissolved oxygen. If that same sample had a salinity value of 9

ppt, then it would hold 8.621 mg/L of dissolved oxygen. therefore, to obtain

accurate mg/L readings, it is important to know the salinity of the water you

will be testing and to input that value into the instrument. the salinity of fresh

water is typically 0-0.5 ppt and seawater is typically 35 ppt. You will also have

the opportunity to enter or modify the Salinity correction value during Do

calibration.

CALIBRAtIoN - DISSoLVED oXYGEN

the Pro Plus offers several options for calibrating dissolved oxygen: Do% in

water saturated air, Do mg/L and Do ppm in a solution of known dissolved

oxygen determined by a Winkler titration, and a Zero point. If performing a

zero point calibration, you must also perform a %, mg/L, or ppm calibration

following the zero calibration. For both ease of use and accuracy, YSI recommends

performing the following 1-point Do % water saturated air calibration:

It is not necessary to calibrate in both % and mg/L or ppm.

Calibrating in % will simultaneously calibrate

mg/L and ppm and vice versa.

Calibrating Do % in Water Saturated Air:

1-Point Calibration

the supplied sensor storage container (a grey sleeve for a single port cable or a

screw on plastic cup for the dual-port and Quatro cables) can be used for Do

calibration purposes.

Moisten the sponge in the storage sleeve or plastic cup with a small amount of

clean water. the sponge should be clean since bacterial growth may consume

oxygen and interfere with the calibration. If using the cup and you no longer

have the sponge, place a small amount of clean water (1/8 inch) in the plastic

storage cup instead.

Make sure there are no water droplets on the Do membrane or temperature

sensor. then install the storage sleeve or cup over the sensors. the storage

sleeve ensures venting to the atmosphere. If using the cup, screw it on the cable

and then disengage one or two threads to ensure atmospheric venting. Make

sure the Do and temperature sensors are not immersed in water. turn the

instrument on and wait approximately 5 to 15 minutes for the storage container

to become completely saturated and to allow the sensors to stabilize.

Press Cal

. Highlight Probe ID or User ID

if you wish to add, select, edit, or delete an ID.

Probe ID must be enabled in the System GLP

menu to appear in the Calibrate menu. User

ID will appear automatically. Select ‘None’

if you do not want a User ID stored with the

calibration. When enabled, these IDs are stored

with each calibration record in the GLP file.

i

Advertising