Troubleshooting, Help error messages – YSI Professional Plus User Manual

Page 39

Advertising
background image

68

69

Long-term Storage - oRP

to store, remove the sensor from the cable and seal the vacant port with the

provided plug. Fill the original shipping/storage vessel (plastic boot or bottle)

with buffer 4 solution and then submerge the sensor into the solution. the sensor

should remain submerged in the solution during the storage period; therefore,

make certain that the vessel is sealed to prevent evaporation and periodically

check the vessel to ensure the sensor does not dry out.

Long-term Storage Temperature: 0 to 30°C (32 to 86°F)

Long-term Storage - Ammonium, Nitrate, and Chloride

the key to ISE sensor storage, short or long-term, is to make certain that the

sensor does not dry out. Sensor junctions that have been allowed to dry out due

to improper storage procedures may be irreparably damaged by the dehydration

and will require replacement. You can attempt to rehydrate the sensor by

soaking it (preferably overnight) in the sensor’s high calibration solution before

attempting to calibrate.

the recommended storage of these sensors is in moist air. Remove the sensor

from the cable and seal the vacant port with the provided plug. Place the sensor

in its original shipping storage vessel (plastic boot or bottle) with a small amount

of tap water or its high calibration standard. the vessel should remain a saturated

air environment. the sensor only needs to be kept in moist air, not submerged.

Make certain that the vessel is sealed to prevent evaporation.

Long-term Storage Temperature: 0 to 30°C (32 to 86°F)

t R o U B L E S H o o t I N G

Illegal Value may appear during alpha/numeric entry on the message line. this

only appears if the values entered do not match the formatting. this will also

appear in GLP security area if the password is incorrect.

If you forget the GLP Security Password please contact YSI tech Support at

[email protected], 800-897-4151, or +1 937 767-7241.

help

During use of the Professional Plus instrument, press Question

from any

screen to view help messages directly on the display.

error MessaGes

If readings for a certain parameter are over range you will see a series of +++++

and if the readings are under range you will see a series of ----- plus the error

message along the bottom of the screen. If you see a series of ????? that will

indicate that a certain parameter can not be calculated. the following are

potential error messages:

Probe temp over range

Probe temp under range

Case temp over range

Case temp under range

pH over range

pH under range

oRP over range

oRP under range

Cl over range

Cl under range

NH4 over range

NH4 under range

No3 over range

No3 under range

Do over range

Do under range

Conductivity over range

Conductivity under range

Barometer over range

Barometer under range

Error messages for the sensors typically indicate a need to properly clean the

sensor. First verify the sensor is properly setup in the Sensor menu, then conduct

the recommended cleaning and attempt to calibrate the sensor. If this does not

work, it may indicate the useful life of the sensor has been reached and may need

to be replaced. You may also contact technical Support to help determine the

next step.