YSI Professional Plus User Manual

Page 26

Advertising
background image

42

43

If a sensor is Enabled that isn’t connected to the

instrument, the display will show an unstable false

reading, ?????, or ----- next to the units.

DISPLAY - oRP

Press Sensor

, highlight Display and press

enter.

Highlight ISE (ORP) and press enter. You will

not be able to Display the sensor unless it is

Enabled in the Sensor Setup menu.


Press enter to enable (

Q

) or disable (

P

) ORP

mV.

AUto StABLE - oRP

Press Sensor

, highlight Auto Stable and

press enter. Highlight ISE (ORP) and press

enter.

Auto Stable indicates when a reading is stable.

Highlight Enabled and/or Audio Enabled

(instrument will beep when the stability is
achieved) and press enter enable (

Q

) or disable

(

P

). When Auto Stable is enabled, AS will blink

next to the parameter until it is stable. once the

parameter is stable, AS will stop blinking.

the Auto Stable Sensitivity can be decreased or

increased. Highlight Sensitivity and use the left and right arrow keys to slide

the bar. the more sensitive you make it (larger black bar) the harder it is to

achieve stability in a changing environment.

the Auto Stable system works by examining the previous 5 readings, computing

the percent change in the data and comparing that change against a % threshold

value. the % threshold value is determined by the Sensitivity bar setting. the

following chart can be used as a guide when setting the Sensitivity bar.

The actual readings displayed during the calibration will

NOT reflect the updated calibration information. These

values will not change until Cal is pressed to finalize

the calibration and to update the instrument.

orp

oRP sensors can be used on 60510-X, 6051020-X, 6051030-X, 6051010-X, and

Quatro cables.

If using a 605103 pH/oRP combination sensor on a 6051020 or 6051030 cable

you can report both pH and oRP by configuring ISE1 as pH and ISE2 as oRP

in the Sensor Setup menu.

the 605103 pH/oRP combination sensor is not recommended for use on a

6051010 or Quatro cable. If used on one of these cable, only pH will be reported

and oRP will not be measured.

SEtUP - oRP

Press Sensor

, highlight Setup, press enter.

Highlight ISE1 if using a 605102 (oRP sensor)

on a 60510, 6051020, or 6051030 cable. Highlight

ISE2 is using a 605103 (pH/oRP sensor) on a

60510, 6051020, or 6051030 cable. If using a

6051010 or Quatro cable, highlight ISE1 if the

oRP sensor is installed in port 1 or highlight

ISE2 if the oRP sensor is installed in port 2 (a

sensor must be installed in port 1 for port 2 to

operate). Press enter.

Enabled allows you to enable or disable the ISE

function and select which ISE sensor is installed

on the cable. Highlight Enabled and press enter
to enable (

Q

) or disable (

P

) the ISE you selected

previously (either ISE1 or ISE2).

After enabling the ISE function, ensure oRP is

selected as the ISE sensor as shown in screen

shot to the left. If necessary, highlight oRP and

press enter to set the selected ISE to oRP.