Barometer – YSI Professional Plus User Manual

Page 20

Advertising
background image

30

31

Highlight Calibration value and press enter to

manually input the sample’s dissolved oxygen

value. Highlight Accept Calibration and press

enter once the temperature and Dissolved

oxygen readings stabilize. or, press Esc

to

cancel the calibration. If User Field 1 or 2 are

enabled in the GLP menu, you will be prompted

to select the fields after selecting Accept

Calibration. After making your selection,
press Cal

to complete the calibration. After

completing the calibration, the message line

will display “Calibrating Channel...” and then

“Saving Configuration...”.

Calibrating in mg/L or ppm as a titration:

2-Point Calibration with Zero Solution

Place the sensor in a solution of zero Do.

A zero Do solution can be made by dissolving approximately 8 - 10 grams of

sodium sulfite (Na

2

So

3

) into 500 mL tap water. Mix the solution thoroughly. It

may take the solution 60 minutes to be oxygen-free.

Press Cal

. Highlight Probe ID or User ID if you wish to add, select, edit, or

delete an ID. Probe ID must be enabled in the System GLP menu to appear in

the Calibrate menu. When enabled, these IDs are stored with each calibration

record in the GLP file.

After selecting the Probe ID and/or User ID if appropriate, highlight Do and

press enter. Highlight Zero and press enter. Wait for the temperature and

Do% values under “Actual Readings” to stabilize, then press enter to Accept

Calibration. If User Field 1 or 2 are enabled, you will be prompted to select the
fields and then Press Cal

to complete the calibration. the screen will then

prompt for a follow-up second point calibration.

Highlight the desired calibration units (mg/L or ppm) and press enter

to continue with the next point. Rinse the sensor of any zero oxygen

solution using clean water. to complete the second calibration point,

follow the steps under Calibrating in mg/L or ppm as a titration:

1-Point Calibration.

BaroMeter

All Professional Plus instruments contain an internal barometer.

DISPLAY - BARoMEtER

Press Sensor

, highlight Display and press enter. Highlight Barometer and

press enter. the measurement unit options are: mmHg, inHg, mBar, PSI, kPa,

or Atm. only one unit can be displayed at a time. Select None if you do not

want to display a barometric pressure reading.

Whether or not you choose to display the barometer reading, the barometric

pressure will still be used for calibrating Do% and for compensating for pressure

changes if Local DO is enabled.

CALIBRAtIoN - BARoMEtER

the barometer in the instrument is calibrated

at the factory. If the barometer requires

calibration, press Cal

. Highlight Probe ID

or User ID if you wish to add, select, edit, or

delete an ID. Probe ID must be enabled in the

System GLP menu to appear in the Calibrate

menu. When enabled, these IDs are stored with

each calibration record in the GLP file.

After selecting the Probe ID and/or User ID

if appropriate, highlight Barometer and press

enter.

Highlight the desired unit and press enter.