46

47

Enabled allows you to enable or disable the ISE

function and select which ISE sensor is installed

on the cable.

Highlight Enabled and press enter to enable (

Q

)

or disable (

P

) the ISE you selected previously

(either ISE1 or ISE2).

After enabling the ISE function, choose the

parameter you want enabled for that ISE. In

this example, NH4 is selected.

Cl - Chloride

NH4 - Ammonium

No3 - Nitrate

If a sensor is Enabled that isn’t connected to the instrument, the

display will show an unstable, false reading next to the units.

DISPLAY - AMMoNIUM, NItRAtE, CHLoRIDE

Press Sensor

, highlight Display, press enter. Highlight ISE2(NH4), press

enter. You will not be able to Display the sensor unless it is Enabled.

Highlight the value you wish to display and
press enter to enable (

Q

) . Ammonium can

be displayed as NH4-N mg/L (Ammonium),

NH3-N (Ammonia) and/or NH4 mV (sensor

signal).

the same steps would be followed to display

nitrate or chloride.

Ammonia is calculated from the pH, salinity, and temperature readings. If a

pH sensor is not in use, the instrument will assume the sample is neutral (pH

7) for the calculation. If a conductivity sensor (Salinity) is not in use, the

instrument will use the salinity correction value entered in the Sensor Menu

for the calculation (see Salinity Correction within the Dissolved oxygen Setup

section of this manual for more information).

AUto StABLE - AMMoNIUM, NItRAtE, CHLoRIDE

Auto Stable indicates when a reading is stable. When Auto Stable is enabled, AS

will blink next to the parameter until it is stable. once the parameter is stable,

AS will stop blinking.

to enable Auto Stable, press Sensor

,

highlight Auto Stable and press enter. Highlight

ISE1 or ISE2 and press enter.

Highlight Enabled and/or Audio Enabled

(instrument will beep when the stability is

achieved) and press enter to confirm. the Auto

Stable Sensitivity can be decreased or increased.

Highlight Sensitivity and use the left and right arrow keys to slide the bar. the

more sensitive you make it (larger black bar) the harder it is to achieve stability

in a changing environment.

the Auto Stable system works by examining the previous 5 readings, computing

the percent change in the data and comparing that change against a % threshold

value. the % threshold value is determined by the Sensitivity bar setting. the

following chart can be used as a guide when setting the Sensitivity bar.

Sensitivity selected by User

% Data Variance Threshold

100 - Most Sensitive, Sensitivity

bar is set to the far right

0.05%

75

0.62525%

50

1.275%

25

1.8875%

0 - Least Sensitive, Sensitivity

bar is set to the far left

2.5%

Within the Auto Stable menu, you can also

choose to Hold All Readings for as many

parameters as you set for Auto Stable. For

instance, if pH and Ammonium have Auto

Stable enabled and Hold All Readings is also

enabled, then the display will hold the readings

once pH and Ammonium have both reached

their Auto Stable settings. You must press the

Esc key to “release” the held display in order to

take subsequent readings. Hold All Readings

must be reactivated after each use!

i