the sky and then twist tying, zip tying, or rubber banding the bulkhead to the

cable. Making a simple curve to the cable without bending or breaking the cable

will allow you to lower the sensor into the aeration tank while the sensor points

skyward so the bubbles are no longer bursting on the membrane surface.

conductIVIty

the conductivity sensor will provide quick readings as long as the entire sensor

is submerged and no air bubbles are trapped in the sensor area. Immerse the

probe into the sample so the sensors are completely submerged and then shake

the probe to release any air bubbles. occasional cleaning of the sensor may be

necessary to maintain accuracy and increase the responsiveness. to clean the

sensor, use the conductivity cleaning brush with a mild detergent.

p

h/orp

pH and oRP readings are typically quick and accurate. However, it may take the

sensors a little longer to stabilize if they become coated or fouled. to improve

the response time of a sensor, follow the cleaning steps in the Maintenance

section of this manual.

aMMonIuM, nItrate, and chlorIde

these sensors may take a little longer to stabilize if the tips are dirty or fouled. If

installed with a pH sensor, always maintain a clean pH sensor for a more rapid

sensor stabilization.

these sensors can only be used in freshwater.

loGGInG data

Log One Sample is already highlighted in Run mode. Press enter to open a

submenu. If Use Site List and or Use Folder List are enabled in the Logging

Setup menu, you will have to option to select these two items before the data

point is logged. If necessary, use the keypad to create a new Site or Folder name.

If Site List and Folder List are disabled in the System menu, you will not see these

options when logging a sample. once the Site and/or Folder name is selected,

highlight Log Now and press Enter. the instrument will confirm that the data

point was successfully logged.

If you would like to log at a specific interval vs. logging one sample at a time
or vice versa, press System

, then highlight Logging and press enter. Select

Continuous Mode and adjust the time Interval if necessary. on the Run screen,

the option to log will change from Log One Sample to Start Logging based on

the time interval entered in the Logging Menu.

During a continuous log, the Start Logging dialog box on the Run screen will

change to Stop Logging. Press Enter to stop continuous logging.

F I L E S A N D S I t E L I S t S

fIle MeMory

to view the file memory, press File

.

the Data Memory shows a percentage

indicating the amount of memory available.

If the file memory is near 0%, files should be

downloaded to a PC and/or deleted to free up

memory.

VIeWInG saVed data

Press File

, highlight View Data and press enter.

Configuring your data view:

Site: will allow you to view data from one

particular site or all sites. Highlight Site, press

enter, and select the site you wish to view data

from or select All Sites to view data from all

sites.

Folder: will allow you to view data from one

particular folder or all folders. Highlight

Folder, press enter, and select the file you wish

to view data from or select All Folders to view

data from all folders.

Begin Date, Begin Time, End Date, and End

Time: will allow you to view data collected

between a specific time period. Highlight the