dIssolVed oxyGen

the dissolved oxygen sensors will use Probe Current (Do uA) and Probe Slope

(%/uA) as part of their GLP file records. the following information indicates

the acceptable values for each of these readings:

Polarographic Do at 25

º

C, 100% saturated air environment at 760 mmHg

Probe Current

1.25 mil PE membrane

Average 6.15 uA (min. 4.31 uA, max. 8.00 uA)

2.0 mil PE membrane

Average 3.38 uA (min. 2.37 uA, max. 4.40 uA)

1 mil teflon® membrane

Average 16.29 uA (min. 11.40 uA, max. 21.18 uA)

Probe Slope

1.25 mil PE membrane

Average 16.26 % sat/uA (min. 12.51 uA, max. 23.23

uA)

2.0 mil PE membrane

Average 29.56 % sat/uA (min. 22.74 uA, max. 42.23

uA)

1 mil teflon® membrane

Average 6.14 % sat/uA (min. 4.72 uA, max. 8.77 uA)

R E S t o R E D E FAU L t

C A L I B R A t I o N VA L U E S

occasionally, the instrument may need to have the factory calibration default

values restored. In order to accomplish this press Calibrate

, highlight

Restore Default Cal and press enter. Highlight the parameter you wish to restore

to default and press enter. Next you will be asked to confirm the operation.

Highlight Yes and press enter to confirm.

A C C E S S o R I E S / PA R t N U M B E R S

Cable Part Number*

Description

6050000

Professional Plus Instrument

60510-1, -4, -10, -20, or -30

1, 4, 10, 20, or 30-meter cable for ISE/temp

60520-1, -4, -10, -20, or -30** 1, 4, 10, 20, or 30-meter cable for Do/temp
60530-1, 4, -10, -20, or -30

1, 4, 10, 20 or 30-meter cable for Cond/temp

6051010-1, 4, -10, -20, or -30

1, 4, 10, 20, or 30-meter cable

for ISE/ISE/temp

6051020-1, -4, -10, -20, or -30 1, 4, 10, 20, or 30-meter cable for ISE/Do/

temp

6051030-1, 4, -10, -20, or -30

1, 4, 10, 20, or 30-meter cable for ISE/Cond/

temp

6052030-1, -4, -10, -20, or -30 1, 4, 10, 20 or 30-meter cable for Do/Cond/

temp

605790-1, -4, -10, -20, or -30

1, 4, 10, 20 or 30-meter Quatro cable for Do/

Cond/temp/ISE/ISE

605107

1-meter pH/temp single junction lab-grade

combo electrode

605177

4-meter pH/temp single junction lab-grade

combo electrode

605108

1-meter oRP/temp single junction lab-grade

combo electrode

605178

4-meter oRP/temp single junction lab-grade

combo electrode

605109

1-meter pH/oRP/temp single junction lab-

grade combo electrode

605179

4-meter pH/oRP/temp single junction lab-

grade combo electrode

Sensor Part Number

Description

605202

Galvanic Do sensor

605203

Polarographic Do sensor

605101

pH (ISE)

605102

oRP (ISE)

605103***

pH/oRP Combination (ISE)

605104****

Ammonium (ISE)

Advertising