Conductivity – YSI Professional Plus User Manual

Page 21

Advertising
background image

32

33

Highlight Calibration Value and press enter to

manually enter the correct “true” barometric

pressure. Next, highlight Accept Calibration,

and press enter. If User Field 1 or 2 are enabled,

you will be prompted to select the fields and
then press Cal

to complete the calibration

or press Esc

to cancel the calibration.

Laboratory barometer readings are usually “true”

(uncorrected) values of air pressure and can be used “as is”

for barometer calibration. Weather service readings are

usually not “true”, i.e., they are corrected to sea level, and

therefore cannot be used until they are “uncorrected”. An

approximate formula for this “uncorrection” is below:

true BP = [Corrected BP] – [2.5 * (Local Altitude in ft. above sea level/100)]

conductIVIty

Conductivity sensors are supplied with 60530-X, 6051030-X, 6052030-X , and

Quatro cables. Conductivity sensors are built into the 60530-X, 6051030-X, and

6052030-X cables and are not replaceable. Conductivity/temperature sensors

are shipped with the Quatro cable, must be installed, and are replaceable.

SEtUP - CoNDUCtIVItY

Press Sensor

, highlight Setup, and press

enter. Highlight Conductivity, press enter.

Enabled allows you to enable or disable the

conductivity measurement. Highlight Enabled
and press enter to activate (

Q

) or deactivate (

P

)

conductivity. Disable conductivity if you do

not have a conductivity sensor connected to the

instrument.

If a sensor is Enabled that isn’t connected to the instrument, the

display will show an unstable, false reading next to the units.

Temp Ref (Temperature Reference) is the reference temperature used for

calculating temperature compensated Specific Conductance. this will be the

temperature all Specific Conductance values are compensated to. the default

is 25 °C. to change the Reference temperature, highlight Temp Ref and press

enter. Use the numeric entry screen to enter a new value between 15.00 and

25.00 °C. Next, highlight <<<ENtER>>> at the bottom of the screen and press

enter on the keypad to confirm.

%/°C (Percent per Degree Celsius) is the temperature coefficient used to

calculate temperature compensated Specific Conductance. the default is

1.91% which is based on KCl standards. to change the temperature coefficient,

highlight %/°C and press enter. Use the numeric entry screen to enter a new

value between 0 and 4%. Next, highlight <<<ENtER>>> at the bottom of the

screen and press Enter on the keypad to confirm.

TDS Constant is a multiplier used to calculate an estimated tDS (total

Dissolved Solids) value from conductivity. the multiplier is used to convert

Specific Conductance in mS/cm to tDS in g/L. the default value is 0.65. this

multiplier is highly dependent on the nature of the ionic species present in the

water sample. to be assured of moderate accuracy for the conversion, you must

determine a multiplier for the water at your sampling site. Use the following

procedure to determine the multiplier for a specific sample:

Determine the specific conductance of a water sample from the site;

1.

Filter a portion of water from the site;

2.

Completely evaporate the water from a carefully measured volume of the

3.

filtered sample to yield a dry solid;

Accurately weigh the remaining solid;

4.

Divide the weight of the solid (in grams) by the volume of water used (in

5.

liters) to yield the tDS value in g/L for this site; Divide the tDS value in g/L

by the specific conductance of the water in mS/cm to yield the conversion

multiplier. Be certain to use the correct units.

If the nature of the ionic species at the site changes between

sampling studies, the TDS values will be in error. TDS cannot

be calculated accurately from specific conductance unless the

make-up of the chemical species in the water remains constant.

to change the multiplier, highlight TDS Constant and press enter. Use the

numeric entry screen to enter a new value between 0 and 0.99. Highlight

<<<ENtER>>> at the bottom of the screen and press Enter on the keypad to

confirm.

i

i

i

Advertising