B&K Precision 4045B - Manual User Manual

Page 85

Advertising
background image

85

FREQ 5E3

FREQ MAXIMUM

FREQ MIN

Query

Syntax: FREQuency?[<ws>MAXimum|MINimum]

Examples: FREQ?

FREQ? MAX

Response: NR3

Considerations:

1) The MIN and MAX arguments refer to currently

settable minimum or maximum.

2) FIXed is alias for CW.

Point Rate

RATE <point rate>

This command is used to set the point rate. It is

coupled with the frequency of the waveform by the

relation:

Frequency = 1/(Point Rate * Wavelength)

Thus changing the point rate will result in a change

in frequency.

Arguments

Type: Numeric

Units: s, ms, us, ns

Range: 20ns to 50s