85
FREQ 5E3
FREQ MAXIMUM
FREQ MIN
Query
Syntax: FREQuency?[<ws>MAXimum|MINimum]
Examples: FREQ?
FREQ? MAX
Response: NR3
Considerations:
1) The MIN and MAX arguments refer to currently
settable minimum or maximum.
2) FIXed is alias for CW.
Point Rate
RATE <point rate>
This command is used to set the point rate. It is
coupled with the frequency of the waveform by the
relation:
Frequency = 1/(Point Rate * Wavelength)
Thus changing the point rate will result in a change
in frequency.
Arguments
Type: Numeric
Units: s, ms, us, ns
Range: 20ns to 50s