4 measurement board, Installation 3.8.4 measurement board – Hach-Lange ORBISPHERE Model 51x User Manual User Manual

Page 40

Advertising
background image

38

Installation

3.8.4 Measurement board

The different measurement boards for the EC and TC sensors are illustrated in

Figure 17

and

Figure 18

below. The type of board is easily identified by the color of the J8 connector. For EC

boards this connector is colored orange, and for TC boards it is colored black.

Note:

The colors indicated are the wire colors in the sensor cable.

NOTICE

It is extremely important that sensors are connected to the correct measurement board.
Connecting a TC sensor to an EC measurement board (and vice versa) will cause irreparable
damage to the measurement board.

Figure 17 EC Measurement board

Figure 18 TC Measurement board

Figure 19 Connector J7

Figure 20 Connector J8

Connector J7 (inputs & outputs)

Measurement alarms relays

1.

Common

2.

Output relay 1

3.

Output relay 2

4.

Output relay 3

Note:

Relays N.O. or N.C. depends on the jumper

position on the relays (see

Measurement alarm

relays on page 39

).

Analog current (or voltage) outputs

5.

GND

6.

Output 1

7.

Output 2

8.

Output 3

Digital inputs

9.

EC sensor: Not used
TC sensor: Hold input (short circuit to pin
12)

10.

Not used

11.

Not used

Advertising