Calibration menu
7.9 Calibration reports
Calibration reports
Once a calibration is completed (for a gas or pressure sensor), the calibration report is updated with the
new details. The calibration report contains data for the last 50 calibrations.
For full details on the data displayed for all the different type of calibration reports, refer to the examples
listed under
.
Each calibration record will contain parameters useful for traceability. For instance, it will contain:
•
the user name and ID
•
the date and time
•
the calibration coefficient
•
all the measurements which influence the calibration (temperature, barometric pressure, current,
etc.)
Calibration reports are generated after a sensor calibration. In addition, once a smart sensor is
connected to the instrument for the first time, a calibration report is automatically generated giving details
of the last calibration of that sensor.
For interfering gas calibration reports, you will be asked to select the interfering gas from those available.