39

Installation

Note:

When changing the type of sensor (e.g. from a 31xxx EC sensor to a 31xxxS smart EC sensor)

please contact your local Hach Lange representative.

3.9 Measurement alarm relays

Analog inputs (when ext. press. sensor is
directly connected)

12.

EC sensor: Not used
TC sensor: GND

13.

Green: Input ext. press. sensor P+

14.

White: Input ext. press. sensor P-

15.

Red: Output ext. press. sensor +

16.

Black: Ground (GND)

Analog inputs (when ext. press. sensor
extension, Part N° 32548.xx, is used)

12.

EC sensor: Not used
TC sensor: GND

13.

Green: Input ext. press. sensor P+

14.

Yellow: Input ext. press. sensor P-

15.

White: Output ext. press. sensor +

16.

Brown: Ground (GND)

Note:

On multi channel systems, the external pressure sensor must be wired to the channel 1

measurement board, but the signal is used to compensate all channels.

Connector J8 (sensor)

Note:

Remember, this connector is colored orange for EC sensors and black for TC sensors.

A1100 EC sensor

31xxx EC sensor

31xxxS smart EC

sensor

TC sensor

Cable

sensor

wall/panel

mount

LEMO 10

sensor

portable

1)
2)
3)
4)
5)
6)
7)
8)
9)

10)

Guard electrode
RS485A+
Thermistor A
Anode electrode
RS485B-
Thermistor B
GND
+ 5V
Cathode electrode
Not used

Guard electrode
Not used
Thermistor A
Anode electrode
Not used
Thermistor B
Not used
Not used
Cathode electrode
Not used

Guard electrode
I2C-SCL
Thermistor A
Anode electrode
I2C-SDA
Thermistor B
GND
+ 5V
Cathode electrode
Not used

GND for power
V2 signal
Solenoid
Relay coil
+12V power
+24V power
V3 signal
GND for signal
-5V power
Temperature

Yellow

Pink

Grey

Red

Purple

White

Black

Green

Blue

Brown

Brown

Red

Orange

Yellow

Green

Blue

Purple

Grey

White

Black

Figure 21 Output relays

The three output relays are located on the measurement
board (see also

Figure 17

and

Figure 18 on page 38

).

They can be individually configured to Normally Open (NO) or
to Normally Closed (NC) by physically moving the jumper on
each relay. The picture example on the left (EC measurement
board):

Upper relay is set to NC

Middle relay is set to NO

Lower relay is shown with no jumper, to show the 3 pins.

Note:

For all measurement boards, J4 is relay 1, J5 is relay 2 and J6