The main board:

Writes the latest measurement data to the Profibus Input Buffer.

Checks if a command written by the Profibus Master must be executed (Profibus Output
Buffer). If a command is to be executed, the instrument executes it and writes the result
(status, data, etc.) in the Profibus Input Buffer.

All numbers are coded in “Big Endian” format, and float values are coded according to IEEE
Standards. The field types “Byte” and “Double Word” are unsigned.

Measurements

Measurements are formatted in the Profibus Input Buffer as follows:

Note:

The table above shows information for a 3-channel instrument. This will be adjusted according to the

number of channels in use.

Name

Type

Size

Offset

Barometric pressure

Input float

32 bits

0

Barometric pressure unit

Input byte

8 bits

4

Channel 1 gas concentration

Input float

32 bits

5

Channel 1 gas unit

Input byte

8 bits

9

Channel 1 temperature

Input float

32 bits

10

Channel 1 temperature unit

Input byte

8 bits

14

Channel 1 external pressure

Input float

32 bits

15

Channel 1 external pressure unit

Input byte

8 bits

19

Channel 1 events

Input double word

32 bits

20

Channel 1 measurement index

Input double word

32 bits

24

Channel 2 gas concentration

Input float

32 bits

28

Channel 2 gas unit

Input byte

8 bits

32

Channel 2 temperature

Input float

32 bits

33

Channel 2 temperature unit

Input byte

8 bits

37

Channel 2 external pressure

Input float

32 bits

38

Channel 2 external pressure unit

Input byte

8 bits

42

Channel 2 events

Input double word

32 bits

43

Channel 2 measurement index

Input double word

32 bits

47

Channel 3 gas concentration

Input float

32 bits

51

Channel 3 gas unit

Input byte

8 bits

55

Channel 3 temperature

Input float

32 bits

56

Channel 3 temperature unit

Input byte

8 bits

60

Channel 3 external pressure

Input float

32 bits

61

Channel 3 external pressure unit

Input byte

8 bits

65

Channel 3 events

Input double word

32 bits

66

Channel 3 measurement index

Input double word

32 bits

70