3 h2 sensor calibration, 3 tc gas sensor calibration, Calibration menu 7.2.3 h – Hach-Lange ORBISPHERE Model 51x User Manual User Manual

Page 65: Sensor calibration

Advertising
background image

63

Calibration menu

7.2.3 H

2

sensor calibration

7.3 TC gas sensor calibration

WARNING

Potential Fire and Explosion Hazard. Handle H

2

gas with great care!

H

2

sensor calibration

The recommended method is at known concentration (Direct Value). The gas can be pure H

2

or a

mixture of H

2

with an inert gas (e.g. a mixture of H

2

/N

2

). The known concentration is entered by the user

in the calibration parameter window. The sensor is in contact with the calibration gas in gas phase at
atmospheric pressure.

NOTICE

Make sure the H

2

concentration used for calibration is within the acceptable range for the

membrane. Hach Lange recommends the following calibration gas/inert gas combinations for
calibrating with the following membranes:

Membrane

Recommended calibration gas

2956A

1% H

2

/ 99% N

2

2952A

10% H

2

/ 90% N

2

2995A

100% H

2

29015A

100% H

2

Calibration of the measured gas

Start

Before initiating a calibration process, the calibration parameters must be
set by pressing on the Modify button. The last calibration parameters are
memorized, so this step can be ignored if the correct parameters are
already set.
Similarly, if only the calibration value has changed, then this can be updated
directly instead of pressing the Modify button.
Note:

When calibration is started, a calibration event is set, and the analog

output and relay are put on hold (if selected below) to avoid unwanted alarms
or potential process problems. This resumes when calibration is finished.

Modify calibration
parameters

Gas Phase

: Select liquid or gas (direct calibration only)

Gas unit type

: Partial, fraction or dissolved (dissolved is for

calibration in a liquid only)

Gas unit

: The list of available units depends on unit type selected

above.

Liquid

: Select as appropriate.

Enter the gas concentration according to the value in the calibration
media.

Hold during calibration

: On by default, this stops any output from the

instrument during the calibration process to avoid sending invalid
information to any connected device.

Automatic calibration stop

: If selected, when the stability criteria is

reached, the calibration process stops automatically.

Interferences enabled

: If selected, this takes into account the influence

of interferences during calibration. By default the same interferences as
during measurement are selected.

Press OK to start calibration

Advertising