Advanced, Measurement menu – Hach-Lange ORBISPHERE Model 51x User Manual User Manual

Page 57

Advertising
background image

55

Measurement Menu

Advanced configuration - EC sensor

Enable pressure sensor

: Check as appropriate. See

External pressure calibration (optional sensor)

on page 65

Enable negative concentration

: Check as appropriate. See

O

3

sensor calibration on page 62

TPO enable (portable instruments in “Sample mode” measurement only)

: Check if TPO

calculation is required.

TPA enable (portable instruments in “Sample mode” measurement only)

: Check if TPA

calculation is required.

If TPA is enabled, enter the TPA K coefficient if different from the displayed value.

Advanced configuration - TC sensor

Enable ext. pressure sensor

: Check as appropriate. See

External pressure calibration (optional

sensor) on page 65

Enable negative concentration

: Check as appropriate.

Continuous purge during thermal cut off

: If thermal cutoff has been enabled (see

Measurement

configuration on page 53

), then check this box to ensure that a continuous purge of the TC sensor

takes place while the measurement session is suspended due to the thermal cutoff temperature
value being exceeded.

Note:

To manually set the TC sensor into a continuous purge mode, press the Continuous Purge button

that is available from the Services - Diagnostic - Channel x - Amplifiers menu. See details in the
section entitled

Amplifiers (TC sensor only) on page 102

.

Offset and slope corrections

: Enable correction as appropriate.

If enabled, the correction values for offset and slope must be entered. These values cannot be
negative.

Liquid to gas factor

: Enable correction as appropriate. If checked, the percentage correction factor

must be entered. This value cannot be negative.

Note:

If you believe you need to enable these corrections, it is advisable to contact a Hach Lange Service

Representative first.

Interferences configuration

These options are available to take into account the influence of some components or gases in the
sample during measurement. All available interference corrections are disabled by default.
The following interference corrections are available:

For oxygen measurements - CO

2

, H

2

S, Chlorine, Salt, H

2

For hydrogen measurements - Temperature, H

2

O, He, O

2

For nitrogen measurements - Temperature, H

2

O, O

2

, H

2

Note:

If you believe you need to enable any of the above, it is advisable to contact a Hach Lange Service

Representative first.

Interferences in oxygen measurement
In some applications, like in the beverage industry, there can be high concentrations of carbon
dioxide in the sample. Hach Lange recommends using the CO

2

interference option if a carbon

dioxide concentration of over 1% in gas phase, or 15 ppm in dissolved phase is present.
In the petroleum industry, the detection of oxygen is sometimes hampered by significant
concentrations of hydrogen sulfide in the sample. Hach Lange recommends using the H

2

S

interference option if the hydrogen sulfide concentration exceeds 0.15% in gas phase, or 5 ppm in
dissolved phase. To operate the oxygen sensor in these conditions requires using a different
sensor and electrolyte.

When using this mode your system will experience sensitivity loss of about

50 times higher than the minimum sensitivity for the membrane.
The instrument can also compensate for the presence of high hydrogen levels in the sample.

CO2 or H2S

: Select CO

2

, H2S or All disabled.

Chlorinity/Salinity

: Select Chlorinity, Salt or All disabled. For chlorinity or salt, it is required to

enter the actual concentration in the sample.

H2

: Select H

2

enabled or disabled. It is required to enter the interfering gas pressure, which is the

partial pressure of H

2

in the sample.

When enabled, the H

2

interference must be calibrated. See

Interference calibration (EC sensor) on

page 64

. A warning pop-up window is displayed to remind the user of this action.

Advertising