Communication Menu
The gas, temperature and barometric pressure unit values are coded as defined in the following
tables:
Note:
If the instrument stops sending measurement data to the module, then after 30 seconds the module
sets the event mask to the value PROFIBUS-DP value not updated (0x80000000) bit mask.
Gas unit
Value
Temperature unit
Value
Barometric
pressure unit
Value
bar
0
K
0
bar
0
mbar
1
°C
1
mbar
1
Pa
2
°F
2
psia
2
kPa
3
atm.
3
hPa
4
Pa
4
psia
5
kPa
5
atm.
6
hPa
6
mbar->bar
9
Pa->KPa
10
%Vbar
12
ppm Vbar
13
%Vext
14
ppm Vext
15
ppm Vbar->%Vbar
16
ppm Vext->%Vext
17
ppm
18
ppb
19
g/l
20
mg/l
21
µg/l
22
%O
2
23
%Air
24
Note:
For the field “Event”, please see the column “Bit mask
.
g/kg
25
V/V
26
%W
27
cc/kg
28
ml/l
29