Communication Menu
9.1.2 Example of use
In this example we use:
•
One PC with a RS232 port.
•
One "RS-485<->RS232 converter"
Procedure:
1.
Connect both RS-485 wires of the instrument to the "RS-485<->RS232 converter".
2.
Connect the "RS-485<->RS232 converter" to the PC RS232 port using a standard cable
(RS232 DB9 straight cable).
On the PC:
1.
Run "Hyperterminal" on the PC.
2.
Configure the PC COM port used (e.g. COM2). Menu "File/Properties/Configure".
3.
Configure the parameters "Baud rate", "Parity", "Nb of stop bits", "Byte size" (Menu
"File/Properties/Configure"). Use the same parameters for the instrument and the PC.
4.
Configure the "Font = Courier 10" (Menu "View/Font").
5.
Connect "Hyperterminal" (Menu "Call/Call").
6.
Save the data received in the file of your choice (Menu "Transfer/Capture Text/Start").
On the instrument:
1.
Use the menu "Communication/RS-485 Simple/Send files" and the button "All files". When
the transfer is finished, close the file with "Hyperterminal" (Menu "Transfer/Capture
Text/Stop"). Now, all the reports are saved in a text file on your PC.
Configuration report example
The "Configuration" below is given for a one channel instrument.
INSTRUMENT CONFIGURATION
Measurement mode . . . . . . . . .Continuous mode
Pressure unit . . . . . . . . . . . . . . .[bar]
Temperature unit. . . . . . . . . . . .[°C]
Storage mode . . . . . . . . . . . . . .Rolling buffer
Storage RAM time. . . . . . . . . . .10 [s]
Storage FLASH time . . . . . . . . .3600 [s]
Channel 1
Membrane . . . . . . . . . . . . . . . . .2956A
Medium . . . . . . . . . . . . . . . . . . .Liquid
Gas unit. . . . . . . . . . . . . . . . . . .mbar
Liquid. . . . . . . . . . . . . . . . . . . . .Water
Resolution displayed . . . . . . . . .3
Thermal cut off . . . . . . . . . . . . .Enabled60.0 [°C]
Alarms
Low Low . . . . . . . . . . . . . . . . . .Disabled0.000000 [mbar]
Low . . . . . . . . . . . . . . . . . . . . . .Disabled0.100000 [mbar]
High. . . . . . . . . . . . . . . . . . . . . .Enabled5.000000 [mbar]
High High . . . . . . . . . . . . . . . . .Disabled1.100000 [mbar]
Hysteresis . . . . . . . . . . . . . . . . .1 [%]
Delay . . . . . . . . . . . . . . . . . . . . .0 [s]
Filter State. . . . . . . . . . . . . . . . .Disabled
Type . . . . . . . . . . . . . . . . . . . . .Median
Depth. . . . . . . . . . . . . . . . . . . . .5
Central depth. . . . . . . . . . . . . . .3
Interferences
CO
2
or H
2
S . . . . . . . . . . . . . . . .All disabled
Chlorinity/Salinity. . . . . . . . . . . .All disabled19.000000 [g/l]
H
2
. . . . . . . . . . . . . . . . . . . . . . .Disabled0.100000 [bar]