LUMIStox 300 Operating Manual

Dr. Bruno Lange GmbH

______________________________________________________________________________

______________________________________________________________________________

Page: 10

2.3 Initial settings for luminescent bacteria tests for determining GL
and EC

Initial settings have to be entered for the luminescent bacteria test to enable the GL or

EC value to be calculated. The necessary settings are requested by the LUMIStox 300

after the evaluation mode <GL> or <EC> has been selected. Their significance and the

method of entering them are explained below.

<N> indicates that all the current settings are to be adopted.

The LUMIStox 300 then switches to the start of the

measurement procedure. If <Y> is chosen, the following

settings are requested in sequence.

The length of the dilution series can be entered with the

numeric keyboard (at least 3 and no more than 9 dilution

stages) or the current setting can be confirmed with <ok>.

Preliminary dilution of a toxic sample is rejected with <no >.

The current setting is adopted with <ok>.

After confirmation with <yes >, the required preliminary dilution

can be entered with the numeric keyboard

(not shown). It is taken into account in the EC or GL value.

Only preliminary dilutions from the dilution series referred to in the DIN standard

are therefore allowed for GL determinations.

settings

change ? J

N

dilution steps

?

4

ok

pre dilution: no
yes no

ok