6 Error messages

concerning the test:

Message

Meaning

Measure

EC20, EC50 not

calculated

The EC20 or EC50 value are not included

within the limits of the dilution series.

If necessary repeat the test with a

new dilution series to get values

“around” 20 resp. 50 % inhibition.

GL not calculated

All inhibition values were above 20%

If necessary repeat the test with a

more diluted sample.

not calc.

Colour correction could not be carried out

for the It values concerned because the

extinction was greater than 1.8.

Repeat the test with a prediluted

sample.

PMT too much light

The photomultiplier received too much

light.

Follow exactly the description of

reactivation and pipetting the

bacteria.

concerning the measuring instrument:

Message

Meaning

Measure

continue, or ESC to

enter EMB

The system diskette is not inserted, or

cannot be read.

Insert the diskette or change it.

LED check

LED is not emitting light

temp error

Temperature outside the monitoring range.

Put the instrument “off” and, after

10 seconds, “on” again. If the error

still occurs you will have to inform

the DR LANGE service (telephone

numbers see p. 1)

menu error

test error

The system diskette is destroyed.

Remove the diskette from the

instrument and insert the second

one, delivered with the instrument

Err: 1001 – Err: 1010

Error in different hardware or software

parts

Put the instrument “off” and, after

10 seconds, “on” again. If the error

still occurs you will have to inform

the DR LANGE service (telephone

numbers see p. 1)

Err: 1000

The instrument has lost the calibration data

Err: 1011

Error in software version

Call the DR LANGE service to

install the necessary data again via

telephone.