LUMIStox 300 Operating Manual

Dr. Bruno Lange GmbH

______________________________________________________________________________

______________________________________________________________________________

Page: 6

1.3 Additional information

You should take account of the following advice to ensure that your LUMIStox 300

works correctly and reliably.

The system diskette must be inserted into the drive before

the instrument is switched on! Whenever the instrument

is moved, the diskette must first be removed from the

drive.

The LUMIStox 300 measuring instrument should not be

operated in an ambient temperature below 16

°C or above 29°C,

otherwise problems may occur with the cooling of the

measuring shaft.

Do not operate the instrument in direct sunlight!

Soiling impairs the functioning of the cuvette lowering system. For

this reason, do not pipette reagents into measuring cuvettes in

the measuring shaft. The measuring shaft should also be closed

in the <exit > mode when the measurements have been

completed. When the LUMIStox 300 is in use, the measuring shaft

is automatically closed after a 10-minute idle period. It can be

opened again by pressing any key.

Before any measurements the LUMIStox 300 should have been

switched on for the least 30 minutes so that the photomultiplier

and the cooled components are ready for operation.