4.2.2 Practical advice: The geometrical dilution series:

Position the necessary number of LUMIStox measuring cuvettes in row A of the

LUMIStherm. A dilution series of the sample as described in the example below

requires 9 cuvettes for the sample dilutions and 1 cuvette for the control solution.

Pipette into the measuring cuvettes, in sequence:

2% NaCl solution

then prepared sample

Use a pipette to transfer,

proceeding from left to right, 1.5

ml into the following cuvette. Mix

each sample dilution by drawing

it into the pipette two or three

times.

There is no need to change the

pipette tip.

The 2% NaCl control solution always stands in A1. The highest sample con-

centration, in this case the undiluted sample, is on the far right, in this case A10.

1,5

1,5

1,5

1,5

1,5

1,5

1,5

1,5

0,0

1.

1,5

1,5

2.

3.

ml

ml

ml

=

1,5

1,5

1,5

1,5

1,5

1,5

1,5

dilution: 1:

1

2

4

8

16

32

64

128

256

1,5