8 the lumistox luminescent bacteria test (lck484) – Hach-Lange LUMISTOX 300 LPV321 User Manual User Manual

Page 21

Advertising
background image

LUMIStox 300 Operating Manual

Dr. Bruno Lange GmbH

______________________________________________________________________________

______________________________________________________________________________

Page: 21

2.8 The LUMIStox luminescent bacteria test (LCK484)

This test is a luminescent bacteria test for operational analysis. One package of

preserved luminescent bacteria enables up to 5 samples to be determined.

Work steps:

1.

Prepare sample(s) and control solution (2% NaCl solution).

2.

Reactivate preserved bacteria as described in package circular.

3.

Pipette 0.5 ml bacteria suspension into measuring cuvettes (1 control and 1 test

solution per sample).

4.

Quickly add 0.5 ml control solution or sample to the bacteria suspensions.

5.

Wait for the incubation period of 15 minutes to elapse.

6.

Select the evaluation mode <test> and the LUMISmini luminescent bacteria test

<484> for LCK484 on the LUMIStox 300.

7.

Measure the control and test solutions and record or print the results.

Sequence of operations on the measuring instrument:

Select test

Measure the first 2% NaCl control solution

Measure the first sample

Record or print result, measure next control

solution and sample, etc.

t e s t O D c h r o n i

r e f e r e n c e c h e c k

p l e a s e w a i t

r e f . c u v e t t e

m e a s u r e

s a m p l e

m e a s u r e

3 2 %

i n h i b .

t e s t

4 8 4

4 8 8

Advertising