4.2 Determining EC and GL values

4.2.1 Preparing a dilution series

Three or more dilution levels of a sample are needed to determine a GL or EC value

with the luminescent bacteria test. A distinction is made between two types of dilution

series.

Geometrical dilution series

The sample is diluted several times by a factor of 2. The following series of sample

dilutions is obtained:

undiluted sample 1:2 1:4 1:8 1:16 etc. diluted sample

During the luminescent bacteria test a volume of the sample is mixed with the same

volume of the suspension of luminescent bacteria. This further dilution of the above

sample dilutions results in the test dilution series:

1:2 1:4 1:8 1:16 1:32 etc. test dilutions

Dilution series referred to in DIN 38412 L34/341

A dilution series as referred to in DIN 38412 contains a combination of two geometrical

dilution series, one of them based on the undiluted sample and the other on a 1:1.5

dilution of the sample. This gives rise to the following series of sample dilutions:

undiluted sample 1:1.5 1:2 1:3 1:4 etc. diluted sample

During the luminescent bacteria test a volume of the sample is mixed with the same

volume of the suspension of luminescent bacteria. This results in a further dilution of

the above sample dilutions results to give the test dilution series. DIN 38412 refers to

the test dilutions of the sample as G-values:

1:2 1:3 1:4 1:6 1:8 etc. G-values