LUMIStox 300 Operating Manual

Dr. Bruno Lange GmbH

______________________________________________________________________________

______________________________________________________________________________

Page: 27

3.1 Colour correction

Light-absorbent colorants in the sample can distort the results of the luminescent

bacteria test. The LUMIStox 300 has an automatic colour and turbidity compensation

feature. This means that the absorbed light can be automatically determined and

corrected with the built-in photometer function each time a measurement procedure is

carried out. The measurement takes a few seconds longer than usual. The colour

correction can be performed in the extinction range from 20 - 1800 mE.

Sequence:

The colour correction feature can be switched off, permanently switched on, or

switched on as required. In the latter case, each time a measurement program is

selected the user is asked whether the colour correction feature should be switched on.

Program selection

Colour correction:

Current setting

perm:

Permanently switched off

aus:

Switched off

var:

Switched on as required

If <var> is set, this query appears each time a measurement

program is selected.

col.corr.: off
perm off var

options

col. relV test

col.corr.: no
yes no

ok