5 printout – Hach-Lange LUMISTOX 300 LPV321 User Manual User Manual
Page 40

LUMIStox 300 Operating Manual
Dr. Bruno Lange GmbH
______________________________________________________________________________
______________________________________________________________________________
Page: 40
5 Printout
An example of a printout of a GL determination with colour correction is shown here.
The listed values for Io and It are the uncorrected original values. The values for
percentage inhibition are colour corrected where necessary. If the system does not
detect any colour or turbidity in a dilution solution (extinction < 0.02), the original values
of It are used to calculate the percentage inhibition. Colour correction cannot be carried
out if the extinction is greater than 1.8.
GL - LCK480
LUMIStox 300 - Version 2.21
date:
25.03.97
timer:
13.58.50
operator:
sample:
bacteria:
comment:
color correction mode
dilution steps:
4
typ of dilution:
DIN
factor of dilution:
1
time of incubation
30 min
test
G-value
intensity
inhib
Io
It
B1
control
1179
1251
1.63
C1
control
1113
1220
-1.63
B2
6
1133
1120
0.121
C2
6
1199
1147
1.32
B3
4
1144
1043
2.46
C3
4
1155
1064
1.93
B4
3
1148
911.0
-1.87
C4
3
1121
879.5
0.78
B5
2
1127
760.4
-2.68
C5
2
1196
786.8
-4.64
fK = 1,08
GL = 2