Operation

Page 18

Manual BÜHLER BU4010ff


Installing the apparatus

- The workings described in this chapter of the manual must only be carried out by qualified
personnel.
- We recommend to install the apparatus as near as possible to the point of extraction. If installed

outside, the device should be fixed on a solid flat base (e.g. plinth). The fixing bars do already

have the necessary fixing holes.

- If the device is equipped with castors (mobile version), please make sure that the two lock-type

castors are locked during operation.

- For better deaeration / ventilation install the apparatus with a distance of approx. 10 cm to the wall.
- The hose has to be laid with a continuous fall from the apparatus to the point of extraction, without

fail. Lower lying points can lead to deposits in the hose which may freeze up in winter.

- To fix the suction hose, we recommend to use the extraction unit available as accessory.
- Fix the hose in such a way that the suction aperture lies in the direction of flow, so that coarse

matter and fibres cannot be forced into the suction aperture.





Remark:

To open the top cover both front screws (left + right)
have to be removed. Then the top cover can be lifted
up to the point where the retention bars snap-in. To
close the cover again, the two retention bars have to
be lifted slightly.

After having closed the cover, the

two screws (left + right) have to be screwed in again.






Mains connection

The device is equipped with a power supply lead.

The connecting data can be found in chapter “Technical Data”.


Connect/disconnect
The device is connected/disconnected by means of the mains plug (except

devices with main switch).


Hose routing

The apertures for the suction hose are on the left and on the right at the front.


Signal inputs

The signal inputs (e.g. for flow meter) are located on the „fold-out“ plate and

can be connected according to the plan of terminal connections shown there.


Messages

The relays required for messages (up to 4) can be fixed on the top hat rail. For

details please see the circuit diagram


Remark

:

The apparatus must be cleaned regularly in accordance with the degree of contamination present. In
view of the quality of samples, we recommend to clean thoroughly especially the wetted parts like
dosing unit, electrodes, distributor, bottles and inlet hose. Failure to do so could result in damage or
destruction to the equipment, device that

are not covered by warranty.