Operation

Page 32

Manual BÜHLER BU4010ff

Troubleshooting – distributor 4010/3010//6010


If a fault develops in your sampler, we can only provide prompt assistance if you give us the type and
serial number of the apparatus concerned. You will find these data on the type plate.
The more precise your description of the fault is, the better our fault diagnosis will be.

Symptom

Possible cause

Action

Chapter


No function


Connections

Water penetrated in distributor
drive


Check connections

Dismantle distributor
head and check for
water penetration.
Contact service
department


Distributor


Program cannot be
started


Wrong distributor has been
entered


Correct entry


Programmin
g


Distributor is
continuously turning


Forked light barrier defective

Pulse generator defective


Contact service
department, return
defective part if
necessary


Distributor


Wrong positioning of
distributor


Interfering pulses.
Pulse generator does not
function correctly


Contact service
department


Distributor



Sample distribution at Bühler 4210



The picture shows the distributor unit:

1. Detailed view of discharge position when rinsing

2. Distributor arm with direct dosing

3. Sample bottles

4. Bottle carrying tray