Operation

Page 22

Manual BÜHLER BU4010ff

itself automatically
proportionally to the flow. This
is achieved by proportionally
scaling the mA signal and the
maximum sample volume. That
means, the sample volume is
automatically adjusted to the
mA
signal.

always proportional to the flow,
that means:
high flow = big sample
volume
low flow = small sample
volume.

Event-
proportional

In this sampling mode the
sample extraction is depending
on an external event signal
(potential-free make contact).
The sample is only extracted as
long as the signal is present.
The sampling interval as well
as the bottle change are
programmed. The bottle is
changed at each new event
signal.
If an event is longer than the
programmed bottle filling time,
two or more bottles will be filled
for this event depending on the
programmed bottle filling time.

Values to program:
Sampling interval
e.g.

00:05 hh:mm


Bottle filling time
e.g.

02:00 hh:mm