Operation

Page 30

Manual BÜHLER BU4010ff

REMARK:

The pinch valve (9) is in inoperative position always closed and will be opened 1x per

hour to avoid sticking of the silicon tube.



Replacement of dosing hose

The silicone hose (BM69301) is only pushed over the lip of the metering vessel. A hose clamp is not
required. Humidify the hose a little bit, fix it at one point and push it over the lip.


Remark:

Conductivity of the sample medium

There are two conductive level electrodes in the metering vessel. Their sensitivity can

be adjusted progressively in the service menu. The factory adjusts the electrodes to

tap water (this adjustment does not have to be changed under normal operating

conditions).


Please note that a wrong adjustment can lead to a flooding of the pneumatic system and to the
damage of component parts.


ATTENTION: Wrong operating conditions can lead to malfunctions or may even damage the

apparatus. No liability or claims under guarantee will be accepted in respect of any

damage arising from the non-observance of these operating instructions.

Troubleshooting – bypass sampling system


If a fault develops in your sampler, we can only provide prompt assistance if you give us the type and
serial number of the apparatus concerned. You will find these data on the type plate.
The more precise your description of the fault is, the better our fault diagnosis will be.

Symptom

Possible cause

Action

Level electrodes do not
react, system is flooded

-Adjustment of conductivity is too
low

-Failure on the power board

-Defective cable connection

-Check adjustment in service
menu

-Exchange power board

-Check cable connection

Sampler does not
extract samples

-Wrong programming

-System is leaking


-No flow

-Inlet hose clogged

-Level electrodes bridged

-Wrong setting of parameters in
the service menu

-Check programming

-Check whether the silicone
hoses are tight

-Check whether there is water

-Remove clogging

-Clean the electrodes

-Check entered values and, if
necessary, reset them to factory-
set values by reinitialization

Metering vessel
overflowing (backflow)

-Outlet hose is clogged or
squeezed

-Pinch valve does not open
completely

-Remove clogging


-Adjust position of pinch valve