Operation

Page 31

Manual BÜHLER BU4010ff

Sample distribution at Bühler 4010/3010//6010

Ÿ

Rotating distributor

- Motor-driven rotating distributor with encoder (pulse generator) which ensures:
- an exact sample discharge position as well as
- an automatic bottle position identification by means of a forked light barrier.

Remark:

For reasons of operating safety, the distributor first always runs over position 1 again
when changing to the next bottle. Thus a correct positioning even after a malfunction
(voltage loss) is always guaranteed.


Remark 2:

When there are no pulses, the output is switched off and thus neither the motor
nor the gear will be damaged if the distributor is blocked.



ATTENTION:

The distributor is driven by a geared motor. The distributor must under no
circumstances be turned by hand! In case of non-observance, the gear will be
damaged!