Installation – Hach-Lange BUEHLER 3010 4010 4210 4410 Manual User Manual

Page 8

Advertising
background image

Installation

Page 8

Manual BÜHLER BU4010ff

*) with opened top

Weight

100 kg with composite container; higher weight when using several
bottles and/or glass bottles


Power supply

230 V 50Hz
optional 115 V 50/60Hz

Power requirement

Approx. 350 VA (with cooling)

Ambient temperature

-20 - +43°C

Sample temperature

0 – 40°C

Standards

Device meets ISO 5667 standard

Wetted materials

PVC, Silicone, PS, PE, EPDM (optional: glass (Duran50) metering
vessel, sinker weight SS304)



Device designation

4210 (like 4010, with the following modifications)

Device type

Fixed site sampler for sampling of faecal matter

Rinsing water inlet
Rinsing water outlet

Union nut R 3/4" (max. 2 bar)
Hose connector 25 mm

Sampling method

Vacuum system 20-500 ml

Sampling modes

Time-related, flow-dependent, event-related, manual sample extraction

Bottle variants

Plastic
1 x 25 L, 4 x 14 L, 4 x 20 L
12 x 1 L, 23 x 1,0 L
Glass
12 x 1,0 L, 23 x 1,0 L

Sample distribution

Direct distribution with additional position for rinsing water. When using
the rinsing function, the customer has to make sure that the rinsing
water can drain off freely

Overall dimensions

1.290 (1.890*) x 690 x 645 mm
for bottle variants: 1x25 L, 4x14 L, 12x1 L
1400 (2015*) x 930 x 850 mm
for bottle variants: 4x20 L, 23x1 L
*) with opened top

Weight

105 kg with composite container; higher weight when using several
bottles and/or glass bottles



Device designation

4410 (like 4010, with the following modifications)

Device type

Self-emptying fixed site sampler for continuous operation without the
need of personnel (monitoring applications)

Emptying/rinsing

Bottles are emptied, rinsed and re-filled on the same bottle position.
The rinsing head is installed above the bottle.

Bottle emptying for sample
collection

According to the device version: manual bottle emptying at 12 and 24-
bottle version or menu-driven emptying at 2 and 4-bottle version

Hose inlet

Hose connection at dosing unit (top)
Union nut R 3/4", hose ID 12 mm

Hose outlet

Hose connection at dosing unit
Silicone hose 12x2, is only slipped over

Rinsing water inlet

Union nut R 3/4" (max. 2 bar)

Rinsing water outlet

Union nut R 1 1/4". When using the rinsing function, the customer has
to make sure that the rinsing water can drain off freely

Bottle variants

Plastic
2 x 10 L, 4 x 5 L, 4 x 10 L, 24 x 2 L (PPH)
Glass
12 x 1,6 L, 24 x 1,0 L, 24 x 2 L

Overall dimensions

(Hxwxd)
• 1.290 (1930*) x 690 x 645 mm

Advertising