Page 7

Manual BÜHLER BU4010ff

Device designation

4010

Housing

Stainless steel with 40 mm insulation,
(material 1.4301/ SS304) / PS / PC (GF10)
Option:
(material 1.4571/ SS316Ti; SS304 EPOXY-coated;
SS316Ti EPOXY-coated)

Thermostatic control

Self-contained cooling / heating with 4 settings, no-frost.
Temperature in sample compartment: 4°C (setting range: 0,0-9,9°C)

Control

Microprocessor control, foil keyboard, graphical display (128*64 Pixel),
back lit

Data memory

Non-volatile data memory: storage of sampling and malfunction data
like sample extractions, bottle changes, messages, external signals

Programming

12 user programs (can be edited freely)

Program start options

Immediately, at a certain time, by an external signal

Program stop options

End of sampling program after one program run, continuous operation
or x-runs

Pause mode

Interruption of program at any time

Overfilling protection

Adjustable from 1–999 samples/bottle

Interval setting

1 min. to 99 h 59 min. in steps of 1 minute

Pulse setting

1 to 9999 pulses/sample

Manual sample extraction

Possible at any time without interrupting the current program run

Program protection

Up to 5 years after voltage loss

Interface

Mini-USB, RS422/485, RS 232

Communication

Optional: TCP/IP RJ45 WEB board, 4-32GB SD/SDHC memorycard
Optional: Modbus or DP PROFIBUS connection

Languages

Multi-language, selectable

Signal inputs

• 2x analogue: 0/4-20 mA,
• 8x digital (flow, event, 1x free programmable)
Optional: expansion 4x digital, 3 free programmable and
8x Analogue Inputs 0-20mA / 0-10 V
Pulse-length min. 60 ms u. switch-level 7-24V, max. working
restistance 500 Ohm, signal-length max. 30 m

Signal outputs / status
messages

Max. 8 digital outputs, 1x collective message
Optional: expansion 8x digital, 5 are free programmable

Sampling method

-Vacuum system 20-350 ml, plastic vessel
-Vacuum system 20-500 ml, optional
-Vacuum flow-proportional 5-350 ml, optional
-Bypass system 20-250 ml optional

Single sample volume
accuracy

< 2,8 % at standard vacuum system

Suction height

Max. 8 m (at 1013 hPa and stagnant medium)

Pumping speed

>0,5 m/s at a suction height up to at least 7,8 m (at 1013 hPa); pump
capacity can be adjusted electronically.

Suction hose

PVC, L=7,5 m, ID=12 mm (ID 16 mm at flow-proportional option)
Max. hose length 30 m

Sampling modes

Time-related, flow-dependent, flow-proportional (option), event-related,
manual sample extraction

Bottle variants

Plastic
1 x 25 L, 1 x 50 L, 2 x 10 L
4 x 6,0 L, 4 x 10 L, 4 x 14 L, 4 x 20 L, 4 x 25 L
12 x 2,9 L, 24 x 1,0 L, 24 x 2,9 L
Glass
12 x 2,0 L, 24 x 1,0 L, 24 x 2 L

Overall dimensions

(Hxwxd)
1.290 (1.890*)x 690 x 645 mm
for bottle variants: 1x25 L, 1x50 L, 2x10 L, 2x22 L, 4x6 L, 4x10 L,
4x14 L, 12x2 L, 12x2,9 L, 24x1 L
1400 (2015*) x 930 x 850 mm
for bottle variants: 4x20 L, 4x25 L, 24x2,9 L, 24x2 L, 36x1 L