Bypass metering vessel
This sampling method is suitable for use in bypass systems with flow-rates from 3 – 20 L/min.

- Patented motor-driven pinch valve which squeezes the dosing hose or the flow-through hose

below the metering vessel.

- No sensor, thus almost no wear.
- Front opening for ease of metering vessel cleaning
- The sample volume can be adjusted between 20 ml and 250 ml by displacing the filling tube


Metering vessel – structure -

1. Sample inlet
2. Hose connection for overflowing water
3. Flange
4. Screw cap
5. Filling level electrodes
6. Metering vessel
7. Adjustable tube – sample volume
8. Screw connection
9. Pinch valve 1 (sample outlet to distributor/bottles)
10. Silicone discharge tube
11. Pinch valve 2 (dosing)
12. Outlet hose





Method of operation
The desired volume can be adjusted between approx. 20 ml and 250 ml by means of the adjustable
tube. To adjust the desired sample volume, loosen the screw connection and displace the filling tube
upwards or downwards.








1. Continous flow

Inoperative Position

2. Pinch valves 1+2 are closed 3. Water is filled up to the

electrodes









6. Inoperative Position



Bypass sampling system


4. Pinch valve 2 is opened.

Water drains off up to the
level of the dosing tube.


5.

Pinch valve 1 is opened
and the sample is filled
into the bottle.