Ransburg, Basic operation, Kv test jumper – Ransburg Vector R Series Cascade 79523 R90 Waterborne User Manual

Page 33

Advertising
background image

Vector R Series Cascade Applicators - Operation

Ransburg

29

Figure 11: Jumper Location -

Overload Activation

BASIC OPERATION

Triggering

High voltage is actuated by pulling the trigger

to start the flow of atomizing and fan control air

through the applicator. When the applicator is

triggered, an air flow switch is activated, the kV

setpoint is displayed on the kV display, the actual

current draw on the µA display and the high volt-

age light illuminates. Under the µA display is a

bar graph meter that illuminates according to the

actual current draw.

Figure 12: µA Bar Graph Display

The green and yellow regions of the bar graph

meter indicate output current is in the optimum

range for maximum transfer efficiency. The red

region of the bar graph indicates high output current

causing decreased transfer efficiency.

The display at the rear of the applicator also

doubles as a microamp bar graph meter when

high voltage is on (see Figure 13). Its function is

similar to that of the control unit bar graph display.

KV TEST JUMPER

To assist in testing and troubleshooting, a jumper

(J8) has been added to the main PC board. By

covering (shorting) both terminals of this jumper,

the high voltage to the spray applicator can be

activated. Thus, for testing and troubleshooting,

high voltage output can be obtained without the

need to trigger air through the spray applicator.

After testing, the jumper must be repositioned so

that it covers only one terminal (open) or the high

voltage will stay on all the time. (See Figure 8 for

location of test jumper J8.)

†

If jumper J8 is left covering (shorting)

both terminals, high voltage will be on when-

ever AC power is turned on.

C A U T I O N

!

AH-06-01.13

Advertising