Ransburg, Routine schedule – Ransburg Vector R Series Cascade 79523 R90 Waterborne User Manual

Page 36

Advertising
background image

Vector R Series Cascade Applicators - Maintenance

Ransburg

32

ROUTINE SCHEDULE

Follow these maintenance steps to extend the

life of the applicator and ensure efficient op-

er-ation:

Several Times Daily

• Turn the control unit power to OFF!

• Inspect the air cap for paint accumulation.

Clean as frequently as necessary with a soft

bristled brush and a suitable solvent.

• Clean all insulating surfaces in the system.

Remove paint accumulation from the exterior

of the applicator and low voltage cable with a

solvent dampened cloth.

Daily (or at start of each shift)

• Verify that ALL solvent safety containers

are grounded!

• Check within 20 feet of the point of operation

(of the applicator) and remove or ground ALL

loose or ungrounded objects.

• Inspect work holders for accumulated coat-

ing materials (and remove such accumula-

tions).

• Check that atomizer assembly is clean and

undamaged.

†

NEVER soak or submerge the electrical

components of the applicator, i.e., barrel,

hook, or cable. Damage and failure may

occur.

C A U T I O N

!

†

NEVER remove the fluid nozzle assem-

bly while paint is in the applicator or paint

may enter into the air passages. Clogged

or restricted air passages will cause poor

atomization and/or electrical shorting. Air

passages that are clogged with conduc-

tive material can lead to excessive current

output levels and consequent low operat-

ing voltage or long-term electrical damage.

Before undertaking any atomizer mainte-

nance procedure, see "Applicator Assem-

bly Cleaning Procedure" in the "Mainte-

nance" section.

The applicator barrel

MUST be tilted front

down to remove the fluid nozzle. Failure to do

so may allow paint to enter the air passages,

thereby reducing airflow and damaging the

applicator barrel/cascade. Applicators may

be flushed in lieu of tilting. However, they

must be either flushed or tilted down during

nozzle removal!

C A U T I O N

!

• Straighten the applicator electrode if neces-

sary.

• Clean the fluid filter, if used.

• Turn the control unit power ON.

• Run a current/voltage output test.

Electrical Current Output Test

1. Turn the paint and/or solvent supply OFF.

†

Standard electrode is "snap back"

spray wire electrode.

NOTE

†

Paint and/or solvent supply must be

turned off during this test - risk of fire or ex-

plosion.

W A R N I N G

!

AH-06-01.13

Advertising