Configuration Mode Passcode

Sets an eight digit Configuration Passcode. Each digit entered can only be set between 0 (zero)
and 3 (three) inclusive.

Pressing the Change Value softkey causes the following screen to display:

Pressing the Prev Digit softkey moves the highlight left to the previous entered digit.

Pressing the Inc Digit or Dec Digit softkeys increments or decrements the selected digit.

Pressing the Next Digit softkey moves the highlight right to the next digit to be entered.

Pressing the Enter softkey selects the values that have been entered and returns to main
configuration menu.

Energy Level: Shock 1, 2, 3

Selects the energy levels used for the 1

st

, 2

nd

, and 3

rd

consecutive shocks that the defibrillator

delivers when:

In semi-automatic or advisory mode and using pads

Ιn manual mode, using external paddles or pads, with Basic Auto Energy Escalation set
to “Yes”

Energy level 1 is the energy level the unit defaults to when the power is turned on with external
paddles or pads connected. For more information about the Basic Auto Energy Escalation
setting, see the “Basic Auto Energy Escalation” section of this manual.

Auto Analyze

This parameter allows you to specify whether the unit automatically repeats ECG rhythm
analyses after the delivery of shocks in the AED and Advisory modes and works in conjunction
with the Stacked Shocks parameter described next. Enabling this parameter also enables the
occurrence of post-shock CPR intervals. When set to No, which is the Manual mode default, the
unit does not perform foreground analysis automatically (with the exception of the first analysis
on power-up in Auto-Defib mode, if the unit is configured to analyze on power-up.)

When set to Yes, and the Stacked Shocks option is set to 3, the unit automatically analyzes the
patient’s ECG rhythm after the first and second shocks in each three-shock sequence. Following
delivery of the final shock, the unit stops automatic re-analysis and either displays a “CHECK
PULSE” prompt, or immediately begins a CPR interval, depending on additional parameter
settings. This function operates only when the Auto Charge in Advisory Mode is set to Yes.

When set to Semi-Auto Only, the unit functions as described above only when it is in AED
Semi-Auto mode. If the unit is changed to Manual mode From AED Semi-Auto mode, the unit
terminates any automatic analysis/re-analysis sequencing. If the Analyze button is pressed, the
unit analyzes only once.

Stacked Shocks

This parameter specifies the number of stacked shocks that must be administered prior to the
activation of a CPR interval and works in conjunction with the Auto Analyze parameter.