12 lead analysis, 12 lead printout, Print 12 lead measurements – ZOLL M Series Defibrillator Rev YC User Manual

Page 20: 12 lead fax format, Print 2 copies 12 lead, Set lead groups, 6\vwhp&rqiljxudwlrq

Advertising
background image

M S

ERIES

C

ONFIGURATION

G

UIDE

16

The Phone Numbers menu displays.

3. Press Next or Prev softkey to scroll to phone number.
4. Press Delete softkey to delete phone number.

12 Lead Analysis

This setting cannot be changed in the current version of software.

12 Lead Printout

When set to “Standard” the 12-lead 4x3 printout will use the standard lead format.

When set to “Cabrera” the 12-lead 4x3 printout will use the Cabrera-style lead format.

Print 12 Lead Measurements

When set to “Yes” the 12SL™ Analysis will produce and print a measurements matrix
including measurements for each lead. Global measurements will be produced and printed
regardless of this setting. This setting affects the strip chart and summary report printouts as
well as the fax transmission. When set to “Yes” any faxed 12-lead report will contain two pages.
The second page of the report will contain the computed measurements matrix.

12 Lead Fax Format

When set to “4x3 Standard” the 12-lead format on the fax output is identical to the format on
the strip chart, with the addition of a 10-second Lead II rhythm strip. (If the 12 LEAD
PRINTOUT configuration option is set to “Cabrera” the fax will be Cabrera as well.)

When set to “2x6” the 12-lead format on the fax output will be configured for two columns of 6
leads, each with 5 seconds of data. The fax will not contain the 12SL interpretative statements
or the 10-second Lead II rhythm strip. In addition, you must set “Auto Transmit After 12 Lead
Analysis” to “Yes.”

The setting “4x3 Median Compl” is supported in software versions higher than 17.00. If “4x3
Median Compl” is selected, the 12-lead ECG waveforms on the strip chart, summary report,
and fax will contain a single median complex beat computed by the GE Medical Systems
Information Technologies 12SL analysis program.

Print 2 Copies 12 Lead

When set to “Yes”, two copies of the 12-lead 4x3 report and 12SL analysis will print on the
strip chart. This setting only affects 12-lead reports generated by pressing the Acquire softkey.

Set Lead Groups

This setting allows the user to configure two custom lead groups as described in the 12-Lead
ECG Monitoring insert (part number 9650-0215-01). Pressing the Change Value softkey will
cause the following screen to appear:

Pressing the Prev Lead softkey will select the previous lead.