19

MUSE Site and Location

Sets the Catalyst MUSE site and location identifiers needed to transmit 12-lead ECG data to a
Catalyst MUSE system. The medical facility’s Catalyst MUSE system administrator provides
these numbers when you initially set up access to the system. The Catalyst MUSE site has a
value between 01 and 32, and the location has a value between 001 and 600.

Note:

You must also configure the Catalyst MUSE communications phone number. See the
“Fax/Communication Phone Numbers” section of this guide.

To set the Catalyst MUSE site and location:

1. From the main System Configuration menu, press Next Item softkey until “MUSE SITE

AND LOCATION” highlights.

2. Press the Change Value softkey to display the MUSE Site and Location menu:
3. Press Prev Digit or Next Digit softkey to select digit and move to previous or next digit.
4. Press Inc Digit or Dec Digit softkey to change value of digit.

Repeat steps 3 and 4 until site has been entered.

5. Press Enter softkey to move cursor to first digit of location.

Repeat steps 3 and 4 until location has been entered.

6. Press Enter softkey to accept site and location values and return to main configuration menu.

12 Lead Secs Per Lead

Sets the number of seconds of ECG data printed for each lead on the 12-lead 4x3 strip. This
setting only affects 12-lead reports generated by pressing the Acquire softkey.

12 Lead Print Speed

Sets the print speed of the 12-lead 4x3 reports to either 25mm/second or 50mm/second. This
setting only affects 12-lead reports generated by pressing the Acquire softkey.

Upload Baud Rate

Sets the upload transmission rate to 9600 bps, 38400 bps, or 115200 bps. This option is only
applicable to units equipped with Bluetooth or RS-232 communications hardware.

Trending Report Zoom Level

Sets the power up resolution for the displayed trend report. Trend data values for all
physiological parameters are simultaneously sampled by the M Series and stored in memory
once each minute.

When displaying this data in tabular format on the M Series screen, however, the user can
choose to view all or only a subset of the stored trend data. The factory default zoom level
causes all one-minute data samples to be displayed when viewing the trend table. If lower
resolution is desired, this configuration option can be set to display only the samples taken
every 5, 10, 15, 30, or 60 minutes. All data samples are printed on the strip chart regardless of
this setting.