Etco, Option nibp option ptca settings option – ZOLL M Series Defibrillator Rev YC User Manual

Page 33

Advertising
background image

Configuration Option Tables

29

EtCO

2

Option

NIBP Option

PTCA Settings Option

Feature

Options

Default

EtCO

2

Options

Set/Review

Set/Review

Default EtCO

2

Units

mmHg, kPa, %

mmHg

Displayed Zoom Level

(in mmHg): 0-12.5, 0-25, 0-50, 0-75, 0-100

(in kPa): 0-1.7, 0-3.3, 0-6.6, 0-10.0, 0-13.3

0-50 mmHg

0-6.6 kPa

Default EtCO

2

Comp Setting

None, O2, N2O, O2 & N2O

None

EtCO

2

Average On Power Up

1 breath, 10 seconds, or 20 seconds

10 seconds

Feature

Options

Default

NIBP Units

mmHg, kPa

mmHg

NIBP Auto Interval Default

2.5, 5, 10, 15, 20, 30, 45, 60, 90, 120 minutes

30 minutes

Trigger NIBP Meas on BP Alarm

Yes/No

Yes

Trigger NIBP Meas on HR Alarm

Yes/No

No

NIBP First Inflate Pressure Default

(in mmHg): 120, 140, 160, 180, 200, 220, 240, 260

(in kPa): 16.0, 18.7, 21.3, 24.0, 26.7, 29.3, 32.0, 34.7

160 mmHg

Feature

Options

Default

Message Enabled

Yes/No

Yes

Threshold

(in mV): 0.8, 0.9, 1.0, 1.1, 1.2, 1.3, 1.4, 1.5, 1.6, 1.7, 1.8,
1.9, 2.0

1.5 mV