Etco, Option nibp option ptca settings option – ZOLL M Series Defibrillator Rev YC User Manual
Page 33

Configuration Option Tables
29
EtCO
2
Option
NIBP Option
PTCA Settings Option
Feature
Options
Default
EtCO
2
Options
Set/Review
Set/Review
Default EtCO
2
Units
mmHg, kPa, %
mmHg
Displayed Zoom Level
(in mmHg): 0-12.5, 0-25, 0-50, 0-75, 0-100
(in kPa): 0-1.7, 0-3.3, 0-6.6, 0-10.0, 0-13.3
0-50 mmHg
0-6.6 kPa
Default EtCO
2
Comp Setting
None, O2, N2O, O2 & N2O
None
EtCO
2
Average On Power Up
1 breath, 10 seconds, or 20 seconds
10 seconds
Feature
Options
Default
NIBP Units
mmHg, kPa
mmHg
NIBP Auto Interval Default
2.5, 5, 10, 15, 20, 30, 45, 60, 90, 120 minutes
30 minutes
Trigger NIBP Meas on BP Alarm
Yes/No
Yes
Trigger NIBP Meas on HR Alarm
Yes/No
No
NIBP First Inflate Pressure Default
(in mmHg): 120, 140, 160, 180, 200, 220, 240, 260
(in kPa): 16.0, 18.7, 21.3, 24.0, 26.7, 29.3, 32.0, 34.7
160 mmHg
Feature
Options
Default
Message Enabled
Yes/No
Yes
Threshold
(in mV): 0.8, 0.9, 1.0, 1.1, 1.2, 1.3, 1.4, 1.5, 1.6, 1.7, 1.8,
1.9, 2.0
1.5 mV