17

Pressing the Next Lead softkey will select the next lead.

Pressing the Change Lead softkey will change the selected lead. Repeatedly pressing this key
will cycle the selected lead through all twelve possibilities.

Pressing the Enter softkey will return to main configuration menu.

12 Lead Freq Response

When set to “0.05-150 Hz 4x3” the 12-Lead Filter Setting defaults to 0.05-150 Hz 4x3.

When set to “0.05-150 Hz Cont.” the 12-Lead Filter Setting defaults to 0.05-150 Hz Cont.

When set to “0.05-40 Hz 4x3” the 12-Lead Filter Setting defaults to 0.05-40 Hz 4x3.

Refer to the 12- Lead ECG Monitoring Operator’s Guide insert (part number 9650-0215-01) for
details on each setting.

Lead Group Default

When set to “Standard” the 12-Lead Group Setting defaults to Standard.

When set to “Custom Group 1” the 12-Lead Group Setting defaults to the setting defined under
Custom Group 1.

When set to “Custom Group 2” the 12-Lead Group Setting defaults to the setting defined under
Custom Group 2.

Refer to the 12- Lead ECG Monitoring Operator’s Guide insert (part number 9650-0215-01) for
details on each setting.

Print 12 Lead Interpretation

When set to “Yes” the 12SL interpretative statements will print following the acquisition of a
12-lead ECG. This setting affects the strip chart and summary report printouts as well as the fax
transmission.

Defib Default to Pads

When set to “Yes” switching the unit to DEFIB mode will set the active lead to Pads/Paddles
regardless of the configuration setting for “Display Lead on Power-Up.”

Cell Phone Type

This setting controls the active cellular phone profile for PCMCIA modem cards. It has no
effect on landline functionality. Pressing the Change Value softkey selects different phone
makes.

Select the phone make corresponding to your target cellular phone. For additional information
on cellular phone compatibility contact The Supply Net, Inc. at

www.thesupplynet.com

or

(800) 826-0279.

EtCO

2

Options

Pressing Set/Review causes a submenu to display, which allows review of or changes to the
following capnography options:

Default EtCO2 Units:

Sets EtCO

2

units to mmHg, kPa, or %

Displayed Zoom Level:

Sets the default scale setting for the displayed capnographic
waveform in the EtCO

2

units selected above. Does not

affect the printed waveform or the waveform data stored on
the data card.

Default EtCO2 Comp Setting: Sets the default compensation for CO

2

to None, O

2

, N

2

O, or

O

2

+N

2

O.

EtCO2 Average on Power Up Sets the time period over which the EtCO

2

values are

averaged to: 1 breath, 10 seconds (default), or 20 seconds.