background image

M S

ERIES

C

ONFIGURATION

G

UIDE

28

Manual Mode

Feature

Options

Default

Auto Generate Strips

Yes/No

Yes

Retain SYNC after Defib

Yes/No

No

Print 3 Leads When Leads Sel.

Yes/No

No

Auto Transmit After 12 Lead

Yes/No

No

Enable Data Relay

Yes/No

No

Fax/Communication Phone Numbers

Set/Review

Set/Review

12 Lead Analysis

Current 10 sec

Current 10 sec

12 Lead Printout

Standard/Cabrera

Standard

Print 12 Lead Measurements

Yes/No

No

12 Lead Fax Format

4x3 Standard, 4x3 Median Compl, 2x6

4x3 Standard

Print 2 Copies 12 Ld Analysis

Yes/No

No

Set Lead Groups

Set/Review

Set/Review

Custom Group 1
(3 leads)

Custom Group 2
(3 leads)

For each lead:

I, II, III, aVR, aVL, aVF, V1, V2, V3, V4,V5,V6

Custom Group 1:
II

III

aVF

Custom Group 2:
II

III

V1

12 Lead Freq Response

0.05 - 150 Hz 4x3, 0.05 - 150 Hz Cont.,
0.05 - 40 Hz 4x3

0.05 - 150 Hz 4x3

Lead Group Default

Standard, Custom Group 1, Custom Group 2

Standard

Print 12 Lead Interpretation

Yes/No

Yes

Defib Default to Pads

Yes/No

No

Cell Phone Type

Motorola, Nokia, NEC, OKI, Sony, Ericsson

Motorola

Basic Auto Energy Escalation

Yes/No

No

MUSE Site and Location

Site: 01 - 32

Location: 001 - 600

Site: 01

Location: 001

12 Lead Secs Per Lead

2.5, 5.0, 7.5, 10.0 seconds

2.5 seconds

12 Lead Print Speed

25 mm/sec, 50 mm/sec

25 mm/sec