Remove card prompt at pwr off, Sync defib excluded, Ptca settings – ZOLL M Series Defibrillator Rev YC User Manual

Page 27: Enable data relay

Advertising
background image

Configuration Settings

23

Remove Card Prompt at Pwr Off

Sets the M Series unit to prompt the user to remove the data card at power-off if the card
contains any information other than self test, and optionally, synchronized cardioversion data.

If set to “Yes”, the unit will display the message “Remove Card” when the user turns the
Selector Switch to the OFF position and the following criteria are met:

1. A data card is installed in the M Series unit.
2. The unit HAS performed a defibrillation, other than a self test, since power-up.

Note:The M Series unit can be configured NOT to display the “Remove Card” prompt if the

card contains only synchronized cardioversion data. See the “Sync Defib Excluded”
option for more information.

3. The unit HAS delivered a pacing current greater than 25 mA since power-up.

When this option is enabled, the “Remove Card” message will be displayed for 10 seconds,
then the M Series unit will shut off.

Sync Defib Excluded

Allows the exclusion of synchronized cardioversion from the criterion of a defibrillation event,
for the purposes of the “Erase Card Prompt at Pwr Off” and “Remove Card Prompt at Pwr Off”
configuration settings.

When set to “Yes”, if the M Series unit has only discharged self test or synchronized
cardioversion defibrillations and has not delivered pacing current greater than 25 mA, the unit
displays the “Erase Card -- Yes/No” prompt at power-off (if so configured). Otherwise, the unit
displays the “Remove Card” prompt at power-off (if so configured).

PTCA Settings

Allows review of, or changes to Percutaneous Transluminal Coronary Angioplasty (PTCA)
settings if the M Series includes the 12-Lead Reperfusion Algorithm option.

With PTCA Settings highlighted and Set/Review selected, press the Change Value softkey to
set the following parameters:

Message Enabled: Determines whether or not the M Series unit prints a PTCA message

(if appropriate).

Threshold:

Sets the threshold, in mV, for issuing the PTCA message. Only
available if the Message Enabled option is set to “Yes”.

Enable Data Relay

Sets the M Series unit to include “DATA RELAY” in the pre-configured locations on the
12 Lead transmission setup screen. To enable this option, the “Auto Transmit After 12 Lead”
option must be set to “Yes”; otherwise, this option is ignored.