Display DO CPR Message

This parameter allows you to specify whether the unit displays the configured CPR prompt
(described below) for the configured CPR time interval after each set of stacked shocks. The
unit behaves differently depending on this particular parameter value.

When set to No (the Manual mode default), the unit behaves as follows:

At the end of any analysis with an outcome of no shock advised, the unit displays

the “NO SHOCK ADVISED” message. The unit then remains idle for the

configured CPR interval duration. After the CPR interval, if the Analyze button has

not been pressed, the “PRESS ANALYZE” message is displayed and remains until

a new analysis is started or the mode is changed.

At the end of a stacked shock sequence (assuming the final shock is delivered) and

when the optional 10 second “CHECK PULSE” message disappears, the CPR

interval begins. The unit remains idle for the duration of the CPR interval. After the

CPR interval, if the Analyze button has not been pressed, the “PRESS

ANALYZE” prompt is displayed and remains until a new analysis is started or the

mode is changed.

If the defibrillator disarms itself after it was charged (because the Defib Ready time

elapsed before a shock was delivered), the unit displays the “PRESS ANALYZE”

prompt until the Analyze button is pressed.

When set to Yes, the unit behaves as follows:

At the end of any analysis with an outcome of no shock advised, the unit displays

the “NO SHOCK ADVISED” message for 10 seconds, after which the message is

cleared and the configured CPR prompt is displayed for the duration of the CPR

interval. After the CPR interval, the “PRESS ANALYZE” prompt is displayed until

a new analysis is started or a mode change occurs. You can start a new analysis

during the CPR interval by pressing the Analyze button.

At the end of a stacked shock sequence (assuming the final shock is delivered) and

when the optional 10 second “CHECK PULSE” message disappears, the CPR

interval begins. The configured CPR prompt is displayed for the duration of the

CPR interval. After the CPR interval, the CPR prompt is cleared and a new analysis

starts automatically.

If the defibrillator disarms itself after it was charged (because the Defib Ready time

elapsed before a shock was delivered), the unit displays the “PRESS ANALYZE”

prompt until the Analyze button is pressed.

The Display DO CPR Message feature is operational only when the Auto Charge in Advisory
Mode and the Auto Analyze options are both set to Yes.

Note:

When set to “Only in Semi-Auto,” the unit functions as described above only when it
is in AED semi-auto mode.

CPR Message After Shock

This parameter determines which CPR message/voice prompt, “IF NO PULSE, PERFORM
CPR,” or “PERFORM CPR,” is issued during the CPR interval after the delivery of the final
shock in a stacked sequence. The selected message is displayed through the entire CPR interval.

CPR Message After No Shock Advised

This parameter determines which CPR message/voice prompt, “IF NO PULSE, PERFORM
CPR,” or “PERFORM CPR,” is issued during the CPR interval after a No Shock Advised
analysis result. The selected message is displayed through the entire CPR interval.

Check Pulse

This parameter works in conjunction with the Auto Analyze parameter to give you the
capability to specify whether the unit displays the prompt “CHECK PULSE” for 10 seconds
immediately after delivery of the final shock in a stacked shock sequence, after each No Shock
Advised analysis result, or both. During this 10-second time interval, The “CHECK PATIENT”
and “PRESS ANALYZE” prompts are suppressed on M Series units in semi-auto mode,