IM 01C21C01-01E

2-3

2. HANDLING CAUTIONS

2) Turn OFF the dielectric strength tester. Then

connect the tester between the shorted SUPPLY
terminals and the grounding terminal. Be sure to
connect the grounding lead of the dielectric strength
tester to the ground terminal.

3) Set the current limit on the dielectric strength tester

to 10 mA, then turn ON the power and gradually
increase the test voltage from ‘0’ to the specified
voltage.

4) When the specified voltage is reached, hold it for

one minute.

5) After completing this test, slowly decrease the

voltage to avoid any voltage surges.

2.9 Installation of Explosion

Protected Type

In this section, further requirements and differences
and for explosionproof type instrument are described.
For explosionproof type instrument, the description in
this chapter is prior to other description in this users
manual.

For the intrinsically safe equipment and explosionproof
equipment, in case the instrument is not restored to its
original condition after any repair or modification
undertaken by the customer, intrinsically safe construc-
tion or explosionproof construction is damaged and
may cause dangerous condition. Please contact
Yokogawa for any repair or modification required to
the instrument.

NOTE

For F

OUNDATION

Fieldbus and PROFIBUS PA

explosion protected type, please refer to IM
01C22T02-01E and IM 01C22T03-00E respec-
tively.

CAUTION

This instrument is tested and certified as intrinsi-
cally safe type or explosionproof type. Please
note that the construction of the instrument,
installation, external wiring, maintenance or
repair is strictly restricted, and non-observance
or negligence of this restriction would result in
dangerous condition.

WARNING

To preserve the safety of explosionproof equip-
ment requires great care during mounting,
wiring, and piping. Safety requirements also
place restrictions on maintenance and repair
activities. Please read the following sections very
carefully.

2.9.1 FM Approval

a. FM Intrinsically Safe Type

Caution for FM intrinsically safe type. (Following
contents refer “DOC. No. IFM012-A12 P.1 and 2.”)

Note 1. Model EJA Series pressure transmitters

with optional code /FS1 are applicable for
use in hazardous locations.

• Applicable Standard: FM3600, FM3610, FM3611,

FM3810, ANSI/NEMA250

• Intrinsically Safe for Class I, Division 1, Groups A,

B, C & D. Class II, Division 1, Groups E, F & G
and Class III, Division 1 Hazardous Locations.

• Nonincendive for Class I, Division 2, Groups A, B,

C & D. Class II, Division 2, Groups E, F & G and
Class III, Division 1 Hazardous Locations.

• Outdoor hazardous locations, NEMA 4X.
• Temperature Class: T4
• Ambient temperature: –40 to 60

°C

Note 2. Entity Parameters

• Intrinsically Safe Apparatus Parameters

[Groups A, B, C, D, E, F and G]
Vmax = 30 V

Ci = 22.5 nF

Imax = 165 mA

Li = 730

µH

Pmax = 0.9 W

* Associated Apparatus Parameters

(FM approved barriers)
Voc

≤ 30 V

Ca > 22.5 nF

Isc

≤ 165 mA

La > 730

µH

Pmax

≤ 0.9W

• Intrinsically Safe Apparatus Parameters

[Groups C, D, E, F and G]
Vmax = 30 V

Ci = 22.5 nF

Imax = 225 mA

Li = 730

µH

Pmax = 0.9 W

* Associated Apparatus Parameters

(FM approved barriers)
Voc

≤ 30 V

Ca > 22.5 nF

Isc

≤ 225 mA

La > 730

µH

Pmax

≤ 0.9 W

• Entity Installation Requirements

Vmax

≥ Voc or Vt, Imax ≥ Isc or It,

Pmax (IS Apparatus)

≥ Pmax (Barrier)

Ca

≥ Ci + Ccable, La ≥ Li + Lcable