Operation, 1 preparation for starting operation, Operation -1 – Yokogawa EJA220A User Manual

Page 28: Preparation for starting operation -1

Advertising
background image

IM 01C21C01-01E

7-1

7. OPERATION

7.

OPERATION

7.1 Preparation for Starting

Operation

The Model EJA210A and EJA220A flange mounted
differential pressure transmitter measures the levels or
densities of liquids. This section describes the opera-
tion procedure for the EJA210A as shown in Figure 7.1
when measuring a liquid level in an open tank.

(a) Confirm that there is no leak in the connecting part

of the transmitter mounting flange.
Remove the plastic dust cap placed in the process
connector (low pressure side).

(b) Turn ON power and connect the BT200.

Open the terminal box cover and connect the
BT200 to the SUPPLY + and – terminals.

(c) Using the BT200, confirm that the transmitter is

operating properly. Check parameter values or
change the setpoints as necessary.
For BT200 operating procedures, see Chapter 8. If
the transmitter is equipped with an integral indica-
tor, its indication can be used to confirm that the
transmitter is operating properly.

Open tank

F0701.EPS

Figure 7.1 Liquid Level Measurement

᭿ Confirming that Transmitter is Operating

Properly

Using the BT200
• If the wiring system is faulty, ‘communication error’

appears on the display.

• If the transmitter is faulty, ‘SELF CHECK ERROR’

appears on the display.

communication error

PARAM
C60:SELF CHECK
ERROR

Communication error
(Faulty wiring)

Self-diagnostic error
(Faulty transmitter)

DATA

DIAG

PRNT

ESC

F0702.EPS

Using the integral indicator
• If the wiring system is faulty, the display stays blank.
• If the transmitter is faulty, an error code will appear

on the display according to the nature of the error.

Self-diagnostic error on
the integral indicator
(Faulty transmitter)

F0703.EPS

NOTE

If any of the error indications above appears on
the display of the integral indicator or BT200,
refer to Subsection 8.5.2 for corrective action.

᭿ Verify and Change Transmitter Parameter

Setting and Values

The following parameters are the minimum settings
required for operation. The transmitter has been
shipped with these parameters. To confirm or change
the values, see Subsection 8.3.3.

• Measuring range ....... See Subsection 8.3.3 (2)
• Operation mode ......... See Subsection 8.3.3 (8)